background image

English

5

MyoSure

®

 XL Tissue Removal Device for Fluent

Recommended Separation Distances

The following lists the recommended separation distances between portable and mobile RF communications equipment and the FLUENT FLUID 
MANAGEMENT SYSTEM. 
Recommended Separation Distances
The FLUENT FLUID MANAGEMENT SYSTEM is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled.  The customer or the user of the FLUENT FLUID MANAGEMENT SYSTEM can help prevent electromagnetic interference by maintaining a 
minimum distance between portable and mobile RF communications equipment (transmitters) and the FLUENT FLUID MANAGEMENT SYSTEM as 
recommended below, according to maximum output power of the communications equipment.

Rated maximum output power of transmitter 
[W]

Separation distance according to frequency of transmitter [m]

150 kHz to 80 MHz 
d = 1.2

P

80 MHz to 800 MHz 
d = 1.2

P

800 MHz to 2.7 GHz 
d = 2.3

P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated by 
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts [W] 
according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption and reflection from structures, 
objects, and people

Tissue Removal Device for Fluent: 50-601XL/50-603XL

The MyoSure XL Tissue Removal Device for Fluent is shown in Figure 1. It is a hand-held 
unit which is connected to the Fluent Fluid Management System via a 6-foot (1.8-meter) 
flexible drive cable and to the Out-FloPak via a 10-foot (3-meter) suction tube. Cutting action 
is activated by a foot pedal. The tissue removal device is a single-use device designed to 
hysteroscopically remove intrauterine tissue.
The flexible drive cable is inserted into the drive cable adapter on the front panel of the Fluent 
Fluid Management System.
The proximal end of the suction tube is connected to the Out-FloPak of the Fluent Fluid 
Management System. 
The suction pressure draws fluid and resected tissue through the tissue removal device’s  
cutting window.

Set-up

The tissue removal device is EtO sterilized. Verify that the tissue removal device is sterile prior 
to use. 
Do not use if the package is opened or damaged. Discard all opened, unused devices.

CAUTION:  

The tissue removal device is intended for single use only. DO NOT RE-
STERILIZE. DO NOT REUSE. Discard tissue removal device after use. Dispose of 
the tissue removal device and packaging according to your facility’s policies 
and procedures concerning biohazardous materials and sharps waste.

WARNING - DANGER: Risk of explosion if used in the presence of flammable 
anesthetics.

1. Review the System Configuration Diagram in Figure 2 for set-up outline.
2.  Refer to the Fluent Fluid Management System Operator’s Manual for instructions on how to set up the Fluent Fluid Management System.
3.  Connect the foot pedal tube to the connector on the front of the Fluent Fluid Management System console.

Figure 1. MyoSure XL Tissue Removal Device 

for Fluent

Figure 2. System Configuration

Summary of Contents for MyoSure XL

Page 1: ...his document contains proprietary data of Hologic Inc No disclosure reproduction or use of any part thereof may be made except by written permission from Hologic REV RELEASE DATE 15 MAY 2019 Artwork c...

Page 2: ...t you should be experienced in hysteroscopic surgery with powered instruments Healthy uterine tissue can be injured by improper use of the tissue removal device Use every available means to avoid such...

Page 3: ...ain distension If this situation occurs tap the foot pedal once or twice to run the tissue removal device the cutting window will then close automatically If it appears that the tissue removal device...

Page 4: ...ay be affected RF emissions CISPR 11 Class A The FLUENT FLUID MANAGEMENT SYSTEM is suitable for use in all establish ments other than domestic and those directly connected to the public low volt age p...

Page 5: ...ended separation distance calcu lated from the equation applicable to the frequency of the transmitter Recommended safety distance d 1 2 P for 150 KHz to 80 MHz d 1 2 P for 80 MHz to 800 MHz d 2 3 P f...

Page 6: ...jects and people Tissue Removal Device for Fluent 50 601XL 50 603XL The MyoSure XL Tissue Removal Device for Fluent is shown in Figure 1 It is a hand held unit which is connected to the Fluent Fluid M...

Page 7: ...s to the set speed and continues until the foot pedal is released 3 Press the foot pedal and observe the tissue removal device action to verify that the motor runs and that the cutting window is close...

Page 8: ...pressly stated in the Agreement i Equipment manufactured by Hologic is warranted to the original Customer to perform substantially in accordance with published product specifications for one 1 year st...

Page 9: ...ox with the returned unit Return used or opened product according to the instructions provided with the Hologic supplied biohazard kit For More Information For technical support or reorder information...

Page 10: ...xide U S federal law restricts this device to sale by or on the order of a physician Do not resterilize Do not use if package is damaged Hologic MyoSure and associated logos are registered trademarks...

Reviews: