MyoSure

®

 XL Tissue Removal Device for Fluent

Immunity test

IEC 60601
test level

Compliance level

Electromagnetic environment - guidance

Voltage dips, short interruptions and 
voltage variations on power supply 
input lines 
IEC 61000-4-11

0% UT (100% dip in the 
UT) for ½ cycle at 0°, 
45°, 90°, 135°, 180°, 
225°, 270°, and 315°

0% UT (100% dip in the 
UT) for ½ cycle at 0°, 
45°, 90°, 135°, 180°, 
225°, 270°, and 315°

Mains power quality should be that of a typical 
commercial or hospital environment.
If the user of the /operator of FLUENT FLUID MAN-
AGEMENT SYSTEM requires continued operation 
during power mains interruptions, it is recom-
mended that the FLUENT FLUID MANAGEMENT 
SYSTEM be powered from an uninterruptible power 
supply or battery.

0% UT (100% dip in the 
UT) for 1 cycle and 70% 
(30% dip in the UT) UT 
for 25/30 cycles at 0°

0% UT (100% dip in the 
UT) for 1 cycle and 70% 
(30% dip in the UT) UT 
for 25/30 cycles at 0°

0% UT (100% dip in the 
UT) for 250/300 cycles

0% UT (100% dip in the 
UT) for 250/300 cycles

Power frequency (50/60 Hz) 
magnetic field  
IEC 61000-4-8

30 A/m

30 A/m

Power frequency magnetic fields should be at 
levels characteristic of a typical location in a typical 
commercial or hospital environment. 

Conducted RF 
IEC 61000-4-6
 
 

Radiated HF interference quantities 
according to IEC 61000-4-3

3 Vrms 
150 kHz to 80 MHz
6 Vrms in ISM Bands 
between 150 kHz and 
80 MHz

3 V/m  
80 MHz to 2.7 GHz

3 Vrms 
150 kHz to 80 MHz
6 Vrms in ISM Bands 
between 150 kHz and 
80 MHz 

3 V/m 
80 MHz to 2.7 GHz

Portable and mobile RF communications equip-
ment used no closer to any part of the FLUENT 
FLUID MANAGEMENT SYSTEM, including cables, 
than the recommended separation distance calcu-
lated from the equation applicable to the frequency 
of the transmitter.  
Recommended safety distance:
d = 1.2√P for 150 KHz to 80 MHz
d = 1.2√P for 80 MHz to 800 MHz
d = 2.3√P for 800 MHz to 2.7 GHz
Where P is the maximum output power rating 
of the transmitter in watts [W] according to 
the transmitter manufacturer, and d is the 
recommended separation distance in meters [m].
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey 

a

should be less than the compliance level in each 
frequency range.

 b

Interference may occur in the vicinity of equipment 
marked with the following symbol:

 

Note *: UT is the AC mains voltage prior to application of the test level.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption and reflection of structures, 
objects, and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, 
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.  To assess the electromagnetic environment due 
to fixed RF transmitters, an electromagnetic site survey should be considered.  If the measured field strength in the location in which the FLU-
ENT FLUID MANAGEMENT SYSTEM is used exceeds the applicable compliance level above, the FLUENT FLUID MANAGEMENT SYSTEM should 
be observed to verify normal operation.  If abnormal performance is observed, additional measures may be necessary, such as re-orienting or 
relocating the changing orientation or the location of the FLUENT FLUID MANAGEMENT SYSTEM.
b) Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

Summary of Contents for MyoSure XL

Page 1: ...his document contains proprietary data of Hologic Inc No disclosure reproduction or use of any part thereof may be made except by written permission from Hologic REV RELEASE DATE 15 MAY 2019 Artwork c...

Page 2: ...t you should be experienced in hysteroscopic surgery with powered instruments Healthy uterine tissue can be injured by improper use of the tissue removal device Use every available means to avoid such...

Page 3: ...ain distension If this situation occurs tap the foot pedal once or twice to run the tissue removal device the cutting window will then close automatically If it appears that the tissue removal device...

Page 4: ...ay be affected RF emissions CISPR 11 Class A The FLUENT FLUID MANAGEMENT SYSTEM is suitable for use in all establish ments other than domestic and those directly connected to the public low volt age p...

Page 5: ...ended separation distance calcu lated from the equation applicable to the frequency of the transmitter Recommended safety distance d 1 2 P for 150 KHz to 80 MHz d 1 2 P for 80 MHz to 800 MHz d 2 3 P f...

Page 6: ...jects and people Tissue Removal Device for Fluent 50 601XL 50 603XL The MyoSure XL Tissue Removal Device for Fluent is shown in Figure 1 It is a hand held unit which is connected to the Fluent Fluid M...

Page 7: ...s to the set speed and continues until the foot pedal is released 3 Press the foot pedal and observe the tissue removal device action to verify that the motor runs and that the cutting window is close...

Page 8: ...pressly stated in the Agreement i Equipment manufactured by Hologic is warranted to the original Customer to perform substantially in accordance with published product specifications for one 1 year st...

Page 9: ...ox with the returned unit Return used or opened product according to the instructions provided with the Hologic supplied biohazard kit For More Information For technical support or reorder information...

Page 10: ...xide U S federal law restricts this device to sale by or on the order of a physician Do not resterilize Do not use if package is damaged Hologic MyoSure and associated logos are registered trademarks...

Reviews: