![Hologic Genius Operator'S Manual Download Page 12](http://html.mh-extra.com/html/hologic/genius/genius_operators-manual_2146529012.webp)
Genius™ Digital Diagnostics System
Instructions for Use
English AW-22963-001 Rev. 001 3-2021 8/13
Table 5. Candida Detection:
Agreement between TIS-assisted Review and OOI Study Results
CAND
TIS
- +
OOI
-
232 5
+
3 17
Agreement
Rates
PPA
77% (57%,
90%)
NPA
99% (96%,
100%)
Detection
Rates
TIS
8.6% (5.7%,
12.6%)
OOI
7.8% (5.1%,
11.7%)
(Diff)
0.8% (-1.8%,
3.4%)
The detection rate for Candida for the Genius Digital Diagnostics System was 7.8%, compared to
8.6% for TIS-assisted review.