Selenia Dimensions and 3Dimensions Systems 

 

Addendum, Radiation Shield Installation 

 

MAN-05397 Revision 002, October, 2017 

Hologic, Inc. 

Page 4 of 4 

 

36 Apple Ridge Road, Danbury, CT 06810 (800) 447-1856 

 

© 2017 Hologic, Inc. The information contained in this document is proprietary. Unauthorized distribution is prohibited.

 

 

Existing Installations (Removal and Replacement) 

1.

 

If not done already: 
a.

 

Power down the system and unplug the Acquisition Workstation. 

b.

 

Remove the front and rear covers of the Acquisition Workstation (see figure 

Removing front, rear, and side covers

 on page 2). 

c.

 

Remove the side covers of the Acquisition Workstation (see figure 

Removing 

front, rear, and side covers

 on page 2). 

2.

 

Loosen the screws (items A - F in figure 

Shield bracket screw fastening

 on page 3) and 

remove the workstation shield bracket. 

3.

 

Remove the front spacer (item 5) and the front reinforcement plate (item 4) as shown 
in figur

Radiation shield installation

 on page 3. The items in steps 4 – 7 refer to this 

same figure. 

4.

 

Carefully remove the current radiation shield (item 3). Two people are required for 
this action. 

5.

 

(If necessary) Remove the back reinforcement plate (item 2) and the back spacer (item 
1) and install replacements. 

6.

 

Lift the replacement shield (item 3) carefully and place in position in front of the back 
reinforcement plate (item 2). Two people are required for this action. 

7.

 

Insert the front reinforcement plate (item 4). Push at the bottom of the front 
reinforcement plate (item 4) to ensure plate is vertical and an equal gap on either 
side, then insert the front spacer (item 5). 

8.

 

Install the shield bracket, hand tightening the screws in the order shown (A through 
F in figure 

Shield bracket screw fastening

 on page 3). After hand tightening screws, 

tighten 1/4 turn or enough to flatten the lock washer. 

 