System Operation
17
EN
500 mmHg using adjustment dial.
·
Use tandem tube (
(12)
Figure 5-5, High vacuum tube) when two canisters
are serially connected to the same vacuum port.
5.4
Turning On the Aquilex System
1. Press the ON/OFF switch. The displays and indicators light up and system
turns on.
2. The system now performs a self-diagnostic test.
3. If a tube set is in the inflow tube holder when the pump switches on, the
pump display
(Fig. 5-1, Front of Pump
(1)
) shows
Remove Tube Set
. The self-
test resumes once the tube set is removed from the roller wheel.
If the system self-test is unsuccessful, the corresponding error messages are
displayed. (See Chapter 9, Error and Warning Messages).
The system has successfully completed the self-diagnostic test when a single au-
dible tone is heard. The message
System OK
is displayed for 5 secs followed by
the message
Insert Tube Set
.
WARNING!
Do not use this system if a defect is suspected or detected during the function
check. This also applies to any obvious defects, especially defects on the power
plug and power cord.
Figure 5-5 High vacuum tube
(6)
Hygiene filter
(7)
Vacuum tube (green connectors)
(8)
High vacuum port (green)
(9)
Canisters
(10)
Tissue trap (MyoSure® procedures)
(11)
Adjustment dial
(12)
Tandem tube
(7)
(8)
(9)
(6)
(10)
(11)
(12)