Hollywog WiTouch Pro User Manual Download Page 1

User Manual

Before Using Your WiTouch Pro Device

Sync the Remote Control and the  

WiTouch Pro Device
1.
  Using the provided screwdriver, remove

the back cover from the WiTouch Pro 

device. 

2.  Remove the clear plastic tab from the

remote control.

3.  On the remote control, press and hold

the (+) and (-) buttons simultaneously

until the LED appears solid green 

(~ 10 seconds). 

4.  Quickly insert the batteries into the 

WiTouch Pro device.

Note:

 The LED on the remote control

will flash several times indicating 

syncing is complete. 

5.  Replace the back cover on the device.

Note:

 You may need to re-sync your  

remote control and WiTouch Pro device 

when you replace the battery in the 

remote control. This will ensure communi-

cation between your remote control and 

WiTouch Pro device. If your remote control 

and WiTouch Pro device ever lose 

communication, please repeat these 

syncing instructions.

Hollywog

2830 Amnicola Hwy

Chattanooga, TN 37406 USA

+1 423 305 7778

[email protected]

www.hollywog.com/witouchpro

® and © 2015 Hollwog, LLC. All rights reserved.

U.S. Patents 8,972,016, D712,052, D701,610,

D716,958. Other U.S. & Foreign Patents Pending

11S.1004J      09/30/2015

2

3

4

5

1

Summary of Contents for WiTouch Pro

Page 1: ...ral times indicating syncing is complete 5 Replace the back cover on the device Note You may need to re sync your remote control and WiTouch Pro device when you replace the battery in the remote control This will ensure communi cation between your remote control and WiTouch Pro device If your remote control and WiTouch Pro device ever lose communication please repeat these syncing instructions Hol...

Page 2: ...lasts more than five days may indicate a severe condition Stop using the device and consult with your physician Use of the device on the following skin conditions may cause a condition to become worse Do not use the device over or in proximity to these skin conditions abnormal skin skin that is not intact uncleaned unhealthy open wounds rashes swollen red infected inflamed areas skin eruptions e g...

Page 3: ... Gel Pads to the WiTouch Pro Device The gel pads are intended for single person use They will last depending on skin type oils and pH levels approximately two to five applications Replace the gel pads when they no longer adhere completely Follow these steps to apply the gel pads 1 Separate two gel pads 2 Remove the blue liner from the side being applied to the electrode area Do not remove the gree...

Page 4: ... a comfortable high sensory sensation Treatment Recommendations You can leave the device in position for multiple treatments during the day It will automatically turn Off after two 2 hours of inactivity It is recommended you wait a minimum of 30 minutes between treatments Conducting a Treatment Always read the safety warnings before conducting a treatment Follow these steps to conduct a treatment ...

Page 5: ...tective liners back on the gel pads Ensure the pads are completely covered Battery Replacement When battery replacement is needed the LED will flash as follows To replace the WiTouch Pro Device Batteries 1 Using the included 0 Phillips head screwdriver remove the cover 2 Remove the old batteries and place the new batteries in the correct polarity 3 Replace the battery compartment cover To replace ...

Page 6: ... Hollywog LLC Company warrants the WiTouch Pro Product is free of defects in materials and workmanship This warranty shall remain in effect for 1 year 12 months from the date of purchase If these Products fail to function during the 1 year warranty period due to a defect in material or workmanship at the Company s Option Company or the selling retailer will replace the respective Product without c...

Page 7: ...cial or hospital environment NOTE UT is the a c mains voltage prior to application of the test level Guidance and manufacturer s declaration electromagnetic emissions The WiTouch and WiTouch Pro are intended for use in the electromagnetic environment specified below The customer or the user of the WiTouch or WiTouch Pro should assure that it is used in such an environment Emissions test Compliance...

Page 8: ...on electromagnetic immunity The WiTouch and WiTouch Pro are intended for use in the electromagnetic environment specified below The customer or the user of the WiTouch or WiTouch Pro should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 M...

Page 9: ... que la synchronisation est complète 5 Replacez le couvercle sur le dos de la téléco mande Veuillez noter qu il se peut que vous aurez besoin de resynchroniser votre télécommande et l appareil WiTouch Pro lorsque vous replacez la pile dans la télécommande De cette façon vous assurerez la communication entre votre télécommande et l appareil WiTouch Pro Si votre télécommande et l appareil WiTouch Pr...

Page 10: ... appareil pourrait signifier une condition grave Cessez d utiliser l appareil et consultez votre médecin L utilisation de l appareil sur la peau atteinte de l une des conditions suivantes peut entraîner l aggravation de la condition Ne pas utiliser l appareil sur ou à proximité de ces conditions cutanées peau anormale peau qui est non intacte non propre ou en pauvre santé plaies ouvertes irritatio...

Page 11: ...l sur l appareil WiTouch Pro Les coussinets de gel sont destinés à un usage unique Ils dureront dépendamment du type de peau des huiles et des niveaux de pH environ deux à cinq applications Remplacez les coussinets de gel lorsqu ils ne se collent plus entièrement Suivez ces étapes pour appliquer les coussinets de gel 1 Séparez deux coussinets de gel 2 Enlevez la pellicule bleue de la surface qui s...

Page 12: ...isser l appareil en place pour multiples traitements au cours de la journée L appareil s éteindra automatiquement après deux 2 heures d inactivité Il est recommandé d attendre un minimum de 30 minutes entre les traitements Effectuer un traitement Toujours lire les avertissements de sécurité avant d effectuer un traitement Suivez les étapes suivantes pour effectuer un traitement 1 Enlevez les pelli...

Page 13: ...rsque le remplacement de la batterie est nécessaire que la LED clignote comme suit De l article Description Pile WiTouch Pro dispositif L indicateur DEL de l appareil WiTouch Pro clignote toutes les deux 2 secondes lorsque l appareil est en marche 2 piles alcalines AAA L indicateur DEL clignote rapidement en jaune pendant le traitement Télécommande L indicateur DEL de la télécommande clignotera en...

Page 14: ... absence de défauts de matériaux ou de main d oeuvre sur le WiTouch Pro Produit Cette garantie restera en vigueur pendant 1 an 12 mois à partir de la date d achat Si ces produits cessent de fonctionner normalement pendant la période de garantie de 1 an en raison de défauts de matériaux ou de main d oeuvre la Société ou le concessionnaire ayant vendu le produit au gré de la société remplacera le Pr...

Page 15: ...ial ou hospitalier NOTE UT est le a c tension de secteur préalable à l application du niveau de test Conseils et Déclaration du Fabricant émissions électromagnétiques Les appareils WiTouch et WiTouch Pro sont destinés à être utilisés dans l environnement électromagnétique spécifié ci dessous Le client ou l utilisateur des WiTouch et WiTouch Pro doit s assurer qu ils sont bien utilisés dans cet env...

Page 16: ... WiTouch et WiTouch Pro sont destinés à être utilisés dans l environnement électromagnétique spécifié ci dessous Le client ou l utilisateur des WiTouch et WiTouch Pro doit s assurer qu ils sont bien utilisés dans cet environnement Test Immunité Niveau Test IEC 60601 Niveau de conformité Environnement électromagnétique Conseils Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz à 8...

Reviews: