User Manual
Version 1.0 16
2
Protection class
IP22
3
Software function
Freeze drying (manual and automatic).
De-icing (manual)
4
Max recipe number
Six(6)
5
Date
History date, history audit, history alarm can be
stored for a maximum of one month
6
Permissions
3
Level permissions.
(administrator, operator, parameter)
7
User management
B. Technical Configuration
SN
Name
MSM
QTY
1
PLC
Siemens CPU-1200
1
2
Touch screen
Siemens TP700 smart Comfort
1
Utility Requirements (Approx.)
SN
Description
Requirements
1
Total power
19Kw @ 480V, 60Hz, 3P, 5W
2
Clean air for aeration system
40L/min @ 0.5~1bar
3
Room environment temperature
≤25
℃
Standard Documentation
SN
Name
QTY
1
Operation Manual (OM)
1
2
Installation Manual (IM)
1
3
Pipes & Instruments Drawing (P&ID)
1
4
Layout Drawing
1
5
Electric Drawing
1