Inclinometer Button
Pressing the INCL button will put the unit in “inclinometer” mode. Refer to Image C below showing unit in
inclinometer mode.
Image C
Using a “two button” approach the right display shows the “live” readings until the Inclinometer cycle switch,
which is a red button located on the side of the microFET3, is pressed. After the inclinometer cycle switch is
pressed the first reading is moved to the left display, and the right display contains the “live” reading. See
Image D below for location of the red inclinometer cycle switch.
Image D
Pressing the inclinometer cycle switch again “freezes” the data in the right display. Then, press the INCL
button. This will subtract the number in the left display from the number in the right display. The final
inclinometer reading result or angle is displayed in the left display.
To begin the cycle of a new inclinometer test, press the INCL button again. The unit will zero out both displays
to begin the cycle again.
Page 8
Summary of Contents for microFET3