Default values for the PMH-2 instrument
The PMH-2 is shipped with default values programmed in non-volatile
memory.
mV MODE
Dosing = UP
UP
Setpoint = 700
700
Dosing Mode = normal dosing
no.d
Low Alarm mV = 200
200
High Alarm mV = 1000
1000
Flow Switch = Normally open
OP
4-20mA signal output mode = normal
nor
4mA = 000mV
000
20mA = 1000mV
1000
pH MODE
Dosing = UP
UP
Setpoint = 7.00
7.00
Dosing Mode = normal dosing
no.d
Temperature Compensation
Manual (+20.0
o
C)
S.oC (+20.0)
Low Alarm pH = 2.00
2.00
High Alarm pH = 10.00
10.00
Flow Switch = Normally open
OP
4-20mA signal output mode = normal
nor
4mA = 0.00pH
0.00
20mA = 14.00pH
14.00
Of course all values can be customised through the configuration setup.
Entered values are stored in non-volatile memory and are not lost through
power failure.
PMH-2 Instruction manual
PAGE 13