Cleaning and disinfection
85
CLEANING THE COARSE FILTER
The coarse
fi
lter must be cleaned once a week. To do this, proceed as
follows:
1. Pull the cover of the air inlet from the device, as shown in Fig. 47.
2. Slide
the
fi
lter cassette out from the cover of the air inlet, as shown
in Fig. 48.
3. Remove the coarse
fi
lter (black) from the
fi
lter cassette.
4. Clean
the
fi
lter with mild soapy water. Do not use any other agents!
5. Rinse
the
fi
lter thoroughly with clear water.
6. Let
the
fi
lter dry completely in the air.
Instead of cleaning the
fi
lter, you can insert a new one or replace the entire
fi
lter cassette with a new one.
The white
fi
ne
fi
lter cannot be cleaned. It should be inspected visually
weekly and replaced monthly or more frequently if heavily soiled. To do
this, proceed as follows:
1. Pull the cover of the air inlet from the device, as shown in Fig. 47.
2. Slide
the
fi
lter cassette out from the cover of the air inlet, as shown
in Fig. 48.
3. Remove the coarse
fi
lter (black) from the
fi
lter cassette.
4. Remove
the
fi
ne
fi
lter and replace it with a new one.
Fig. 47: Pulling out the air inlet cover
Fig. 48: Sliding out the
fi
lter cassette