Hobo U14 User Manual Download Page 9

HOBO U14 Data Logger User Manual 

9

 

LCD Indicators 

 

Memory

 

Display

 

States

 

State

 

MEM

 

LEFT

 

Hash

 

marks

 

Segments

 

Illustration

 

Stopped

 

On

 

None

 

 

Launched/

 

Logging

 

On

 

Number

 

of

 

segments

 

depends

 

on

 

memory

 

remaining.

 

Highest

 

segment

 

flashes.

 

 

Launch

 

Delaying

 

Flashing

 

All

 

On/Flashing

 

 

Memory

 

Full

 

Flashing

  

(until

 

data

 

is

 

downloaded)

 

All

 

Off

 

 

 

 

External

 

Sensor

 

Removed

 

If

 

an

 

external

 

sensor

 

is

 

disconnected

 

from

 

the

 

logger,

 

the

 

logger

 

will

 

record

 

values

 

of

 ‐

888.88°C/

888.88

 

°F

 

for

 

temperature

 

and

 

0%

 

RH.

 

No

 

External

 

Sensor

 

 

If

 

the

 

U14

002

 

logger

 

is

 

launched

 

and

 

no

 

external

 

sensor

 

has

 

been

 

connected,

 

the

 

display

 

will

 

show

 

dashes

 

‐‐‐

 

and

 

will

 

log

 

false

 

values

 

(

888.88°C/

888.88°F

 

for

 

temperature

 

and

 ‐

888.9%

 

for

 

RH).

 

Temp

 

Only

 

Sensor

 

Installed

 

 

If

 

a

 

temperature

only

 

sensor

 

is

 

connected,

 

no

 

RH

 

value

 

will

 

be

 

displayed.

 

No

 

RH

 

Sensor

 

Installed

 

but

 

Channel

 

Active

 

 

A

 

display

 

of

 ‐‐

%

 

RH

 

indicates

 

that

 

the

 

humidity

 

channel

 

is

 

active

 

in

 

the

 

software

 

and

 

should

 

be

 

turned

 

off

 

to

 

maximize

 

memory

 

space.

 

Otherwise,

 

the

 

unit

 

will

 

record

 

a

 ‐

888.9%

 

RH

 

value

 

at

 

every

 

logging

 

interval

 

Summary of Contents for U14

Page 1: ...es in typical conditions Built in relay for use with Remote Alarm or Automatic Dialer Direct USB interface for launching and data readout using HOBOware Sample logging interval can be set from one sec...

Page 2: ...e 0 100 RH 20 to 50 C 4 to 122 F Accuracy 2 5 from 10 to 90 RH typical to a maximum of 3 5 including hysteresis see Plot B Resolution 0 03 RH 25 C 77 F Response time 2 minutes to 90 in airflow of 1 m...

Page 3: ...1 may require replacement if it is damaged mechanically or by exposure to condensation or chemicals The replacement sensor part number HUM RHPCB 3 or RHPCB 3A is available from Onset or an Onset Autho...

Page 4: ...f the case c Feed the wires through the holes in the relay terminal d Tighten the screws on the terminal e Replace the battery compartment cover f Connect the wires to the Remote Alarm or Automatic Di...

Page 5: ...as never been connected to the computer before it may take a few seconds for the new hardware to be detected Windows only 5 Start HOBOware and open the Launch window From the Device menu select Launch...

Page 6: ...ted when the Host has relaunched the logger or when the sensor reading has returned within limits e Set Relay Contact If you are using an external device that will be triggered by an alarm such as a R...

Page 7: ...re software on the computer 5 Optional If you have more than one logger connected to the computer connect to the HOBO U14 by selecting Device Select and then select the HOBO U14 from the Select Device...

Page 8: ...segments decreases as remaining memory decreases NOTE If the memory is full the logger will no longer log The logger will continue to monitor conditions however and will trigger an alert if the readin...

Page 9: ...will record values of 888 88 C 888 88 F for temperature and 0 RH No External Sensor If the U14 002 logger is launched and no external sensor has been connected the display will show dashes and will l...

Page 10: ...t is damaged mechanically or by exposure to condensation or chemicals The replacement sensor Part No HUM RHPCB 3A is available from Onset or an Onset Authorized Dealer for U14 001 loggers with serial...

Reviews: