Hobo U14 User Manual Download Page 5

HOBO U14 Data Logger User Manual 

5

 

 

HOBO

 

U14

 

with

 

Onset’s

 

ARA

 

Remote

 

Alarm

 

and

 

AVD

45

 

Automatic

 

Dialer

  

NOTE:

 

The

 

polarity

 

must

 

be

 

as

 

shown.

 

3.

 

If

 

you

 

have

 

an

 

External

 

Sensor

 

model

 

(U14

002)

:

 

Plug

 

the

 

external

 

sensor

 

into

 

the

 

RJ

12

 

jack

 

on

 

the

 

side

 

of

 

the

 

HOBO

 

U14.

 

Hold

 

the

 

black

 

rubber

 

flap

 

down

 

against

 

the

 

connector

 

so

 

it

 

slides

 

into

 

the

 

port.

 

NOTE

:

 

The

 

sensor

 

readings

 

will

 

not

 

be

 

displayed

 

until

 

the

 

logger

 

is

 

launched

 

(immediate

 

or

 

delayed)

 

or

 

the

 

status

 

is

 

checked

 

using

 

either

 

HOBOware

 

or

 

the

 

U

Shuttle.

 

The

 

readings

 

will

 

not

 

be

 

logged

 

until

 

you

 

launch

 

the

 

logger

 

later

 

in

 

this

 

procedure.

 

4.

 

Connect

 

the

 

U14

 

to

 

a

 

computer

 

running

 

HOBOware.

 

Plug

 

the

 

large

 

end

 

of

 

the

 

USB

 

cable

 

supplied

 

with

 

HOBOware

 

into

  

a

 

USB

 

port

 

on

 

the

 

computer

 

and

 

plug

 

the

 

small

 

end

 

of

 

the

 

cable

 

into

 

the

 

side

 

of

 

the

 

HOBO

 

U14.

 

If

 

the

 

logger

 

has

 

never

 

been

 

connected

 

to

 

the

 

computer

 

before,

 

it

 

may

 

take

 

a

 

few

 

seconds

 

for

 

the

 

new

 

hardware

 

to

 

be

 

detected

 

(Windows

 

only).

 

5.

 

Start

 

HOBOware

 

and

 

open

 

the

 

Launch

 

window.

 

From

 

the

 

Device

 

menu,

 

select

 

Launch.

 

Configure

 

Launch

 

parameters,

 

including

 

a

 

Launch

 

description,

 

logging

 

interval,

 

and

 

start

 

time.

 

Click

 

Help

 

in

 

HOBOware

 

for

 

details

 

on

 

Launch

 

parameters.

 

6.

 

Configure

 

Alarms

 

(optional).

 

Use

 

the

 

Alarm

 

feature

 

to

 

flash

 

a

 

warning

 

on

 

the

 

LCD

 

if

 

the

 

monitored

 

temperature

 

or

 

relative

 

humidity

 

falls

 

outside

 

user

selectable

 

limits.

 

If

 

the

 

temperature

 

or

 

RH

 

measured

 

by

 

the

 

logger

 

falls

 

outside

 

the

 

selected

 

high

 

or

 

low

 

limit,

 

the

 

word

 

‘Alert,’

 

an

 

alert

 

symbol,

 

and

 

the

 

temperature

 

or

 

RH

 

measurement

 

that

 

went

 

out

 

of

 

limits

 

will

 

flash.

 

7.

 

If

 

you

 

require

 

remote

 

or

 

secondary

 

notification

 

of

 

an

 

alarm

 

condition,

 

you

 

can

 

connect

 

a

 

remote

 

audio

 

alarm

 

and/or

 

an

 

Automatic

 

Dialer

 

to

 

the

 

contact

 

closure

 

as

 

described

 

in

 

Step

 

2.

 

To

 

access

 

the

 

Alarm

 

feature,

 

click

 

the

 

Alarms

 

button

 

in

 

the

 

Launch

 

window,

 

which

 

opens

 

the

 

following

 

dialog

 

box.

 

 

Summary of Contents for U14

Page 1: ...es in typical conditions Built in relay for use with Remote Alarm or Automatic Dialer Direct USB interface for launching and data readout using HOBOware Sample logging interval can be set from one sec...

Page 2: ...e 0 100 RH 20 to 50 C 4 to 122 F Accuracy 2 5 from 10 to 90 RH typical to a maximum of 3 5 including hysteresis see Plot B Resolution 0 03 RH 25 C 77 F Response time 2 minutes to 90 in airflow of 1 m...

Page 3: ...1 may require replacement if it is damaged mechanically or by exposure to condensation or chemicals The replacement sensor part number HUM RHPCB 3 or RHPCB 3A is available from Onset or an Onset Autho...

Page 4: ...f the case c Feed the wires through the holes in the relay terminal d Tighten the screws on the terminal e Replace the battery compartment cover f Connect the wires to the Remote Alarm or Automatic Di...

Page 5: ...as never been connected to the computer before it may take a few seconds for the new hardware to be detected Windows only 5 Start HOBOware and open the Launch window From the Device menu select Launch...

Page 6: ...ted when the Host has relaunched the logger or when the sensor reading has returned within limits e Set Relay Contact If you are using an external device that will be triggered by an alarm such as a R...

Page 7: ...re software on the computer 5 Optional If you have more than one logger connected to the computer connect to the HOBO U14 by selecting Device Select and then select the HOBO U14 from the Select Device...

Page 8: ...segments decreases as remaining memory decreases NOTE If the memory is full the logger will no longer log The logger will continue to monitor conditions however and will trigger an alert if the readin...

Page 9: ...will record values of 888 88 C 888 88 F for temperature and 0 RH No External Sensor If the U14 002 logger is launched and no external sensor has been connected the display will show dashes and will l...

Page 10: ...t is damaged mechanically or by exposure to condensation or chemicals The replacement sensor Part No HUM RHPCB 3A is available from Onset or an Onset Authorized Dealer for U14 001 loggers with serial...

Reviews: