background image

HOBO MX pH and Temperature Logger (MX2501) Manual 

9

burst logging began at 9:06. Burst logging then continued 
until 9:12 when the sensor reading fell back below the 
high limit. Now back in normal mode, the next logging 
interval will be 10 minutes from the last burst logging 
point, or 9:22 in this case. If burst logging had not 
occurred, the next data point would have been at 9:15. 

A New Interval event is created each time the logger
enters or exits burst logging mode. See 

Logger Events

 for

details on plotting and viewing the event. In addition, if
the logger is stopped with a button push while in burst
logging mode, then a New Interval event is automatically
logged and the burst condition is cleared, even if the
actual high or low condition has not cleared.

Statistics Logging 

During fixed interval logging, the logger records data for 
enabled sensors and/or selected statistics at the logging 
interval selected. Statistics are calculated at a sampling rate you 
specify with the results for the sampling period recorded at 
each logging interval. The following statistics can be logged for 
each sensor: 

The maximum, or highest, sampled value, 

The minimum, or lowest, sampled value,

An average of all sampled values, and 

The standard deviation from the average for all sampled
values. 

For example, a logger is configured with the logging interval set 
to 5 minutes and Fixed Interval Logging and Normal selected for 
the logging mode. In addition, all four statistics are enabled 
with a statistics sampling interval of 30 seconds. Once logging 
begins, the logger will measure and record the actual 
temperature and pH sensor values every 5 minutes. In addition, 
the logger will take a temperature and pH sample every 30 
seconds and temporarily store them in memory. The logger will 
then calculate the maximum, minimum, average, and standard 
deviation using the samples gathered over the previous 5-
minute period and log the resulting values. When reading out 
the logger, this would result in 10 data series (not including the 
mV series): two sensor series (with temperature and pH data 
logged every 5 minutes) plus eight maximum, minimum, 
average, and standard deviation series (four for temperature 
and four for pH with values calculated and logged every 5 
minutes based on the 30-second sampling). 

To log statistics: 

1. Tap the HOBOs icon and tap the logger to connect to it. If

the logger was configured with Bluetooth Always Off
enabled, push the switch on the logger to wake it up. If the
logger was configured with Bluetooth Off Water Detect and
is currently underwater, remove it from the water.

2. Once connected, tap Configure. 

3. Tap Logging Mode and then select Fixed Interval Logging. 

4. Select Normal to record the current reading for each

enabled sensor at the logging interval shown at the top of
the screen. Do not select this if you only want to log
statistics.

5. Select the statistics you want the logger to record at each

logging interval: Maximum, Minimum, Average, and
Standard Deviation (average is automatically enabled when
selecting Standard Deviation). Statistics will be logged for all
enabled sensors. In addition, the more statistics you record,
the shorter the logger duration and the more memory is
required.

6. Tap Statistics Sampling Interval and select the rate to use

for calculating statistics. The rate must be less than, and a
factor of, the logging interval. For example, if the logging
interval is 1 minute and you select 5 seconds for the
sampling rate, then the logger will take 12 sample readings
between each logging interval (one sample every 5 seconds
for a minute) and use the 12 samples to record the resulting
statistics at each 1-minute logging interval. Note that the
faster the sampling rate, the greater the impact on battery
life. Because measurements are being taken at the statistics
sampling interval throughout the deployment, the battery
usage is similar to what it would be if you had selected this
rate for the normal logging interval.

7. Tap Done. 

8. Tap Done again to exit the Logging Mode screen. 

9. Tap Start in the Configure screen to load the statistics

settings onto the logger if you are ready to start. 

Setting a Password 

You can create an encrypted password for the logger that will 
be required if another phone or tablet attempts to connect to 
it. This is recommended to ensure that a deployed logger is not 
mistakenly stopped or purposely altered by others. This 
password uses a proprietary encryption algorithm that changes 
with every connection. 

To set a password: 

1. Tap the HOBOs icon and connect to the logger. 

2. Tap Logger Password.

3. Type a password up to 10 characters.

4. Tap Save. 

Only the phone or tablet used to set the password can then 
connect to the logger without entering a password; all other 
mobile devices will be required to enter the password. For 
example, if you set the password for the logger with your tablet 
and then try to connect to the device later with your phone, 
you will be required to enter the password on the phone but 
not with your tablet. Similarly, if others attempt to connect to 
the logger with different devices, then they would also be 
required to enter the password. To reset a password, connect 
to the logger, tap Set Logger Passkey, and select Reset to 
Factory Default or press the button on the logger for 10 
seconds. 

Reading Out the Logger 

To download data from the logger: 

1. Open HOBOmobile.

2. Tap the HOBOs icon and tap the logger to connect to it. If

the logger was configured with Bluetooth Always Off, press

Summary of Contents for MX series

Page 1: ...ately 30 5 m 100 ft line of sight in air Wireless Data Standard Bluetooth Low Energy Bluetooth Smart Logging Rate 1 second to 18 hours Logging Modes Fixed interval normal statistics or burst Memory Modes Wrap when full or stop when full Start Modes Immediate push button date time or next interval Stop Modes When memory is full push button date time or after a set logging period Time Accuracy 1 min...

Page 2: ...cap can detect the presence of water This allows you to configure the logger in a power saving mode in which Bluetooth advertising is active only when the logger is removed from water See Configuring the Logger for details Note The logger will check for the presence of water every 15 seconds when the Bluetooth Off Water Detect power saving mode is selected pH Electrode This is the electrode that i...

Page 3: ...ut from the logger without twisting it 2 If this is the first time you re installing a pH electrode skip to step 3 If a pH electrode is currently installed gently remove it from the sensor end cap Use a foam swab dipped in water to clean the internal O rings Use the foam swab to spread a small dot of silicone based grease along both the O rings inside the sensor end cap Do not use a cotton swab as...

Page 4: ... Configure Select logger settings and load them onto the logger to start logging See Configuring the Logger Readout Download logger data See Reading Out the Logger pH Calibration Calibrate the logger following the instructions in HOBOmobile The logger should be calibrated after replacing or cleaning the electrode See Calibrating the Logger Full Status Details Check the battery level and view the c...

Page 5: ... as prompted making sure the sensor end cap temperature sensor and closure cap are submerged in the solution Follow the instructions on the screen Once the pH reading is stable tap Confirm Buffer 10 If you selected three point calibration repeat steps 8 9 11 Tap Save Calibration 12 Rinse the pH sensor with deionized or distilled water again If you will not be deploying the logger immediately attac...

Page 6: ...e Setting up Alarms for more details 7 Tap Start Logging and select when logging will begin Now Logging will begin immediately after tapping Start in the Configure screen On Next Logging Interval Logging will begin at the next even interval as determined by the selected logging interval On Button Push Logging will begin once you press the switch on the logger for 3 seconds On Date Time Logging wil...

Page 7: ...the High Alarm if you want an alarm to trip when the sensor reading rises above the high alarm value Drag the slider to the reading that will trip the alarm or tap the value field and type a specific reading In the example an alarm will trip when the temperature rises above 85 F 5 Enable the Low Alarm if you want an alarm to trip when the sensor reading falls below the low alarm value Drag the sli...

Page 8: ...ensor Limits 5 Enable High Limit if you want burst logging to occur when the sensor reading rises above a specific reading Drag the slider to the reading that will trigger burst logging or tap the value field and type a specific reading In this example the logger will switch to burst logging when the temperature rises above 85 F 6 Enable Low Limit if you want burst logging to occur when the sensor...

Page 9: ...n select Fixed Interval Logging 4 Select Normal to record the current reading for each enabled sensor at the logging interval shown at the top of the screen Do not select this if you only want to log statistics 5 Select the statistics you want the logger to record at each logging interval Maximum Minimum Average and Standard Deviation average is automatically enabled when selecting Standard Deviat...

Page 10: ...ed because the reading was outside the alarm limits or back within range Note Although the reading may have returned to a normal range during logging an alarm cleared event will not be logged if the logger was set up to maintain alarms until reconfigured New Interval The logger has switched to logging at the burst logging rate or back to the normal rate Water Detect The logger has been placed in w...

Page 11: ...s on the sensor end cap shown above Using your finger spread a small dot of grease along both outer O rings Battery Information The logger requires one user replaceable AA 1 5 V alkaline or lithium battery With a logging interval of 1 minute and at 25 C 77 F battery life is 1 year with Bluetooth Always On selected 2 years with Bluetooth Off Water Detect selected and 3 years with Bluetooth Always O...

Page 12: ...nto position 9 Screw the closure cap back on the logger WARNING Do not cut open incinerate heat above 85 C 185 F or recharge the lithium battery The battery may explode if the logger is exposed to extreme heat or conditions that could damage or destroy the battery case Do not dispose of the logger or battery in fire Do not expose the contents of the battery to water Dispose of the battery accordin...

Page 13: ...dustry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Avis de conformité pour l Industrie Canada Le présent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts ...

Reviews: