23
r=000
Stops the rotational speed, e.g. if shaking should be stopped but tempering be
continued.
Other inputs like e.g. r=0 are interpreted as rotational speed and cause the
motor to rotate with the minimum rotational speed.
Every command must be confirmed with "CR"
In case of unknown commands and/or in case of spelling errors, nothing is
done. The inputs in the setup menu like colour, contrast and brightness of the
display, language, rotation direction etc. can only be done at the operating
panel of the system itself.
Troubleshooting
The display remains blank
Please check if the mains switch on the back side is switched on.
If it is switched on but voltage is present at the socket, check the micro-fuse
and replace if required. (IEC 127-2/III, 250 V, 2 A time-lag).
This fuse - and a spare fuse - are located in the IEC-bushing (in which the con-
nection cable is plugged). The fuse box can be pulled out with a screwdriver.
The system does not cool, heat or shake as set
Check whether the display indicates the correct system type name at power-
on. If not, please contact your local dealer or the HLC by DITABIS service de-
partment.
There are excessive temperature fluctuations
Check the seat of the changeable block by pulling it up. If it is too loose or if
the block exhibits roughness, e.g. contamination, the heat / coolness is not
transferred correctly.