RSC
-
-15000…+15000;
-15000…+15000
Read analog setpoint and analog control value
VONVA [mV]:
analog input setpoint readback
VC [mV]:
analog controller output
RRS
-
0,1
Read analog input source impedance
0: 0
Ω
output impedance of signal source
1: 50
Ω
output impedance of signal source
Default: 1
RSS
-
0,1(, 2)
Read
setpoint
source
0: analog setpoint VONVA (analog input signal on IN)
1: digital setpoint VONVD
2: internal ARB/DDS signal generator (if implemented)
RV1
-
VO_NOM…
VO_NOM
Set static output voltage setpoint
VONVD [V]: digital setpoint VONVD; Default: 0
RVM
-
-VO_NOM*1.5
…VO_NOM*1.5
Read output voltage momentary value
VOAV [V]:
output voltage, actual value
SCD
0…1000
X Set PID-controller D component
Param: PIDD_SP (non-dimensional); Default: 10
PIDD_SP=0 disables the PID D controller component
SCI
0…7;
0,1
X Set PID-controller I component and I limit
Param: PIDI_SP (non-dimensional); Default: 1
PIDI_L; Default: 1
PIDI_SP=0 disables the PID I controller component
PIDI_L=1 limits the control range of the PID I controller component
SCP
1…1000
X Set PID-controller P component
Param: PIDP_SP (non-dimensional); Default: 10
SCR
SRL_MIN…SRL_MAX;
0,1
X Set slew-rate limiter rate and enable
Param: SRLR_SP [V/
μ
s]; Default: 30 #
SRL_ENA; Default: 1 #
SF1
8, 16, 32, 64, 128, 256,
512, 1024
X Set high voltage measurement averaging time
TAVG1 [ms]; Default: 128
Used on VHVPAV_AVG, VHVNAV_AVG
SHS
0,1
Switch
HV-Source
0: OFF Command
1: ON Command
SL1
0…IO_LIM_MAX
X Set amplifier output stage current limit
IO_LIM [mA]; Default: IO_NOM
SL2
0…IHVP_LIM_MAX;
8, 16, 32, 64, 128, 256,
512, 1024
X Set HV source positive current shutdown limit and time
IHVP_SDN_LIM [mA]; Default: IHVP_LIM_NOM
IHVP_SDN_TIM [ms]; Default: 128
SL3
0…IHVN_LIM_MAX;
8, 16, 32, 64, 128, 256,
512, 1024
X Set HV source negative current shutdown limit and time
IHVN_SDN_LIM [mA]; Default: IHVN_LIM_NOM
IHVN_SDN_TIM [ms]; Default: 128
SRR
-
Shutdown reset request
Reset shutdown latches
SRS
0,1
X Select analog input source impedance
0: 0
Ω
output impedance of signal source
1: 50
Ω
output impedance of signal source
Default: 1
SSS
0,1(,2)
X Select setpoint source
0: analog setpoint VONVA (analog input signal on IN)
1: digital setpoint VONVD
2: internal ARB/DDS signal generator (if implemented)
Default: 0
SV1
-VO_NOM…VO_NOM
Set static output voltage setpoint
VONVD [V]: digital setpoint VONVD; Default: 0