background image

 

16 

hivolt.de

 GmbH & Co. KG 

Oehleckerring

 

40

 

  

  

 

D-22419

 

Hamburg

 

  

  

 

Germany

  

 

 

  +49 40 537122-0

 

  

  

 

  +49 40 537122-99

 

  

  

 

[email protected]

 

  

  

 

www.hivolt.de 

5

 

Operation 

5.1

 

Initial Check 

Once the product is delivered, please check the packaging and the device for possible transport 
damage. Please check the device taken out of the packaging for any mechanical defects before the 
unit is put into operation. 
If the device has any signs of damage caused by transport, please immediately inform the shipping 
company so that damages can be claimed. 
 
 

5.2

 

Warning Notices 

-

 

For safe operation of this device it should be put into operation by a qualified electrician according 
to this Operating Manual. 

-

 

The device may only be operated as a component of an overall setup that fully complies with the 
regulations for working with high voltage systems. 

-

 

Output connectors may only be touched when the device is disconnected from the power supply! 
Otherwise, there is a risk of electric shock.  

-

 

The test setup must be fully wired and protected against any contact before the device is put into 
operation. 

-

 

The test setup must be checked each time before the device is put into operation to ensure that it 
is not potentially dangerous. It should be checked that the high voltage connections are faultless 
and the insulation of the wires is not damaged.  

-

 

The high-voltage areas must be blocked in accordance with regulations or secured by other 
means. 

-

 

Once the test setup is connected, any existing capacitances can be charged to high voltage. They 
may carry dangerous voltages even after the device is switched off. 

-

 

The ground bolt on the rear panel of the device must be connected to the central grounding point 
of the test setup and to protective earth. Local regulations on grounding must be observed. 

-

 

No wires or similar objects may project into the device through the ventilation slots 

-

 

If it is suspected that safe operation is no longer possible, the device has to be taken out of 
operation and secured against unintentional operation. 

 

 

This symbol on the output terminals warns of the risk of electric shock. 
 

 

 
 

5.3

 

Temperature Compensation 

To avoid condensation within the device, it should be allowed to reach the room temperature. Please 
unpack the product at least two hours prior to power-up. 
 
 

Summary of Contents for HA61-3B30

Page 1: ...1 hivolt de GmbH Co KG Oehleckerring 40 D 22419 Hamburg Germany 49 40 537122 0 49 40 537122 99 info hivolt de www hivolt de HA61 3B30 High Voltage Amplifier 3000V Operating Manual...

Page 2: ...nitor 15 4 2 Ambient Conditions 15 4 3 Mechanical Specifications 15 5 OPERATION 16 5 1 Initial Check 16 5 2 Warning Notices 16 5 3 Temperature Compensation 16 5 4 Installation 17 5 5 Mains Connection...

Page 3: ...49 40 537122 0 49 40 537122 99 info hivolt de www hivolt de 6 9 Status Word 3 26 7 SIGNAL GENERATOR 27 8 SCPI PROTOCOL 27 9 OPERATION MAINTENANCE 27 9 1 Troubleshooting 27 9 2 Maintenance 28 9 3 Clea...

Page 4: ...t must be secured against unintentional or accidental operation The safety ground must always be connected The ground screw on the rear panel must be connected to the grounding point of the setup and...

Page 5: ...u keiner Gef hrdung kommen kann Die Hochspannung f hrenden Bereiche m ssen vorschriftengerecht abgesperrt oder anderweitig gesichert sein Sollte der Verdacht bestehen dass das Ger t besch digt ist ode...

Page 6: ...put and one 24V compatible INH input are provided to quickly switch the output on and off The maximum output current can be limited to programmable values and programmable behavior When the output cur...

Page 7: ...noise as well as fully arbitrary wave shapes Parameters like waveform amplitude frequency duty cycle rise fall time offset DC value current limit and further functional parameters are controllable vi...

Page 8: ...es The monitor outputs are buffered output impedance 1k and short circuit proof They are able to drive capacitive loads coaxial cables but are not designed to drive cables with a low impedance termina...

Page 9: ...These parameters can be set independently Limiting the instantaneous value of the output current will not result in a shutdown condition but it has an influence on the dynamic behavior of the output...

Page 10: ...r then with a slight overshoot Large capacitive loads driven by high frequency signals may result in a DC offset superimposed on the output voltage This occurs if the final value of the desired output...

Page 11: ...via the 24V control signals INTERLOCK ON and INTERLOCK OFF on the interlock interface or via software command SHS The high voltage can only be switched on when the Interlock shutdown state is cleared...

Page 12: ...on as the interlock circuits are closed IL1 and IL2 at H level This operating mode involves a safety risk STS SDR Shutdown Status Shutdown Reset When an error occurs interlock overload overtemperature...

Page 13: ...olt de Operation without using ON OFF control SDR STS OFF GND INH 2 1 2 1 EMERGENCY STOP X2 INTERLOCK 1 2 3 4 5 6 7 8 9 10 SDR 24V_IL GND STS IL1 ON IL2 INH STS_HV STS_HV The device must be disconnect...

Page 14: ...nge 3000V 3000V Load current range static 30mA 30mA Load current range dynamic t 0 6s 50mA 50mA Power bandwidth CL 50pF VO 6000VPP THD 1 DC 15kHz DC 1 6kHz Bandwidth CL 50pF VI 2VPP 3dB DC 50kHz DC 16...

Page 15: ...re Operation Storage and Transportation 0 25 40 70 C C Relative humidity Operation Storage and Transportation Not condensing 5 5 80 95 Depending on the ambient temperature modulation amplitude and loa...

Page 16: ...e the device is put into operation The test setup must be checked each time before the device is put into operation to ensure that it is not potentially dangerous It should be checked that the high vo...

Page 17: ...se check the available mains voltage The device may be damaged if operated with the wrong mains voltage 5 6 Grounding The M4 grounding nut and screw on the rear panel of the device must be connected t...

Page 18: ...l light up 10 Push the ON push button Warning Now High Voltage can be present at the output connector 11 The HV OK LED will light up the OFF push button and the HV OFF palm button will light up 12 Pus...

Page 19: ...load the drivers from https www ftdichip com Drivers VCP htm and install then manually In the Device Manager entries are displayed under USB Controller USB Serial Converter and under Ports COM LPT USB...

Page 20: ...m pars cr lf Response err id pars cr lf Start character id 1AN Sequence id an alphanumeric character not the id is returned with the response for identification of the command sequence com 3AN Command...

Page 21: ...Read aux values 1 Supply Voltages P24V mV actual value of 24V supply P15V mV actual value of 15V supply N15V mV actual value of 15V supply P48V_HV mV actual value of 48V supply HV P15V_HV mV actual va...

Page 22: ...IHVNAV_AVG A positive HV supply output current averaged value RHM 0 VO_NOM 1 5 VO_NOM 1 5 0 0 IO_M IO_M 0 Read high voltage source momentary values VHVPAV V positive HV supply voltage actual value VHV...

Page 23: ...f the PID I controller component SCP 1 1000 X Set PID controller P component Param PIDP_SP non dimensional Default 10 SCR SRL_MIN SRL_MAX 0 1 X Set slew rate limiter rate and enable Param SRLR_SP V s...

Page 24: ...IO_M mA 150 30 75 75 150 IO measurement range SRL_MIN V s 1 S R Limiter min setting SRL_MAX V s 400 30 S R Limiter max setting IO_LIM_MAX mA 60 30 50 50 100 PA max current limit IHVP_LIM_MAX mA 30 10...

Page 25: ...W_OTP_SDN_HV 4 1 STW_OVP_SDN_HV X 1 OVP Shutdown latched HV OVP_SDN_HV 5 2 STW_OSC_SDN X 1 Oscillation Detect Shutdown STW_OSC_DET 6 4 STW_PWFAIL_SDN X 1 Supply Voltages Error Shutdown STW_PWFAIL_HV S...

Page 26: ...HV I48V Shutdown latched HV 8 1 9 2 STW_OTP_SDN_PA X 1 Overtemperature PA OTP_SDN_PA 10 4 STW_OTP_SDN_HV X 1 HV OTP Shutdown latched HV OTP_SDN_HV 11 8 12 1 STW_OSC_DET X 1 Oscillation Detected OSC_D...

Page 27: ...ot attempt to locate any faults within the device This can be dangerous to life due to the high voltage used in the device In such a case please return the device to the manufacturer after consultatio...

Page 28: ...ated dust has to be removed by blowing out the unit cautiously The manufacturer specified lifetime of the fans is 50000h After about 6 years of continuous operation the fans might be replaced Type SUN...

Page 29: ...SB port to the USB COM interface on HA61 2 open the terminal program and connect to the COM port as shown in the Windows device manager configuration 115 200 kBd no parity bit 8 data bits 1 stop bit 1...

Page 30: ...on the terminal program re connect to the appropriate COM port 14 enter command RID to check the connection and see the new software version number 15 start the update of the HV generator software by...

Page 31: ...ves Low Voltage Directive 2014 35 EU EMC Directive 2014 30 EU and comply with the following European standards EN 61010 1 2010 Safety requirements for electrical equipment for measurement control and...