M-96RFS Setting and Adjustment
■ Initialization Setting Method
□ After installing it on the tank, set it to the initial state for quick response.
■ Tuning Setting
□ Set the default capacitance value in the tank.
No
Function
Method
1
Frequency Tuning
■ Tune the cap trimmer for setting the default value.
■ Adjust clockwise until
DET. LED turns off.
2
If the sensing LED does not turn off
(High capacitance in the tank)
■ Adjust SEN. VR slightly clockwise and then readjust the cap trimmer.
SEN. : 0
Relay Delay Time: 0.5s
Relay Return Time: 0.5s
Fail Safe Mode (NC)
Cap Trimmer: Tuning
[Before tuning the cap trimmer]
[After tuning the cap trimmer]