Maintenance
Cleaning Chamber
At end of a measure and after validating the result, you can easily clean the bowl to remove
the tested sample.
1. Remove the bowl cover;
2. Add additional carbon-black (rubbish or bad quality carbon-black);
3. Start the motor and the application switch automatically in Cleaning mode, and wait
time ends. If you want, you can stop the cleaning process by opening
the front door or by pushing the
Stop
button;
4. Un-tighten the Knurled nuts by turning a bit and take it out by sliding out;
5. Slide the bowl out from its position;
6. Remove the amount of sample to a trash box;
7. Clean the internal surface of the bowl and the drive blades with some cleaning paper
till there is no carbon-black;
8. Set the bowl in place as before.