AP 720S

TM

 

LCD Panel Guide

 

 
 
 
 
 

OPTIGEN® 

Semi-Automated Instrument

 

 
 
 
 
 

 

 

   

 
 
 
 
 
 
 
 

 
 
 
 
 
 

Summary of Contents for OPTIGEN AP 720S

Page 1: ...AP 720STM LCD Panel Guide OPTIGEN Semi Automated Instrument ...

Page 2: ... and information contained herein constitute confidential information of Hitachi Chemical Diagnostics and customer shall not disclose or transfer any of these materials or information to any third party Copyright 2007 Hitachi Chemical Diagnostics Inc 630 Clyde Court Mountain View California 94043 _________________ U S and Foreign Patents Pending All rights reserved No part of this book may be repr...

Page 3: ...hod 2 1 Initialization 4 2 2 Main Menu 5 2 3 Panel Type Selection 6 2 4 Auto Run Process 1 to 7 7 2 5 Auto Run Photoreagent Setup and Process 8 9 2 6 Auto Run Process End 10 2 7 Maintenance 11 2 8 Pause 13 2 9 Start End Process and Position Setup 16 2 10 Alarm 17 3 Flowchart 21 ...

Page 4: ... executable Press the Operation Key to execute the function See Figure 1 Numerical Value Input Key Numerical value can be registered by ENT key If the value entered is out of the defined range beep sounds will be on for 3 times ESC key is used for returning to the original value See Figure 2 Figure 1 Figure 2 ...

Page 5: ... 3 will display approximately 5 seconds after the power is turned on Then the screen will change automatically to the initialization screen Figure 4 After the completion of initialization there is a beep sound and Main Menu screen Figure 5 will display Operation Key Figure 4 Pause To Figure 27 Pause screen ...

Page 6: ...gure 5 is the Main Menu screen after completion of Figure 4 Panel 36 is the default panel Figure 6 is the Main Menu screen after selecting Panel 20 in Figure 7 Operation Key Panel 36 To Figure 7 Panel Type Selection screen Next To Figure 8 Auto Run screen Maintenance To Figure 19 Maintenance screen ...

Page 7: ...isplay and Operation Method 2 3 Panel Type Selection Screen Display Figure 7 Summary Select 36 allergen or 20 allergen panel in Figure 7 Operation Key 36 To Figure 5 Main Menu screen 20 To Figure 6 Main Menu screen ...

Page 8: ...7 2 Screen Display and Operation Method 2 4 Auto Run Process 1 to Process 7 Screen Display Figure 8 Figure 9 Figure 10 Figure 11 Figure 12 Figure 13 Figure 14 Continued on next page ...

Page 9: ...ss 2 or 5 is in operation See Chapter 6 Operation in the AP 720S Instruction Manual o Figure 13 shows incubation process Process 3 or 6 is in operation o Figure 14 is a warning screen during incubation It indicates the door is open with a beep sound After the door is closed the screen will return to Figure 13 In the case the door is open during the operation i e initialization priming rack pettes ...

Page 10: ...6 Figure 16 requests confirmation of Photoreagent setup No key will return to the previous screen Yes key will shift the screen to Figure 17 and continue Auto Run Please note that before pressing Yes key the Operator needs to fill the Photoreagent Reservoir with Photoreagent mixture and load into Sample Reagent Rack See instructions in AP 720S Instruction Manual Operation Key Figure 15 Alarm Off S...

Page 11: ...f the completion of Auto Run The display shows the elapsed time from the completion of Auto Run Alarm Off key will stop the alarm End key will return to the Main Menu screen Figure 5 Rinse key will shift the screen to Figure 22 Operation Key Figure 18 Alarm Off Stop beep sound End To Figure 5 Main Menu screen Rinse To Figure 22 screen ...

Page 12: ...11 2 Screen Display and Operation Method 2 7 Maintenance Screen Display Figure 19 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 Figure 25 Figure 26 Continued on next page ...

Page 13: ...ll start rinsing operation Ok To Figure 23 Rinse Operation screen Back To Figure 19 Maintenance screen Figure 24 Prior to Drain Operation the Operator needs to pull out the tubes from Wash Buffer Bottle and Deionized Water Bottle Ok key will start draining operation Ok To Figure 25 Drain Operation screen End To Figure 19 Maintenance screen Figure 26 Internal clock of the machine can be set in this...

Page 14: ...13 2 Screen Display and Operation Method 2 8 Pause Screen Display Figure 27 Figure 28 Figure 29 Figure 30 Figure 31 Figure 32 Figure 33 Figure 34 Continued on next page ...

Page 15: ... Figure 34 Figure 34 Yes To Figure 3 No To previous screen Supplement At the time of operation failure the instrument will automatically pause and various warning message will be displayed on the screen Even though operation can resume if causes of failure remain the instrument will pause again Please call Service Representative for additional assistance The following are examples of alarm message...

Page 16: ...L C4 L indicates failure of reagent rack rotation carrier drive or sensor failure After you confirm the status press CONTINUE key and check the status o Example Information C4 H C4 H indicates failure of reagent rack rotation carrier drive or sensor failure After you confirm the status press CONTINUE key and check the status o Example Information C5 L C5 L indicates failure of aspiration syringe d...

Page 17: ...f Auto Run so that the operation will be continued without missing steps Start Process Range 1 8 a smaller number than End Process End Process Range 1 8 a greater number than Start Process Position Pette start position range 1 20 Operation Key Figure 35 Start Process key To Figure 2 numerical value input screen with the title Start Process End Process key To Figure 2 numerical value input screen w...

Page 18: ... selection to Retry or End the process Figure 36 Alarm screen of power interruption confirmation Figure 37 Alarm screen of data initialization confirmation Figure 38 Alarm screen of no Sample Reagent Rack selection Figure 39 Alarm screen of no pettes selection Figure 40 Alarm screen of wrong pette setup selection Figure 41 Alarm screen of wrong start pette setup selection Figure 36 Alarm Screen of...

Page 19: ...ted Re entry of program and internal parameters may be required If this is the case please contact your Service Representative Figure 38 Alarm Screen of No Sample Reagent Rack Screen Display Figure 38 Summary If the control software detects that there is no Sample Reagent Rack at the start of the run Figure 38 will appear If the run needs to be continued setup the Sample Reagent Rack and press Ret...

Page 20: ...arm Screen of Wrong Pette Setup Screen Display Figure 40 Summary If the control software detects that the pette setup is wrong at the start of the run Figure 40 will appear If the run needs to be continued setup pettes correctly and press Retry key to continue If the run needs to be cancelled press End key to end the Auto Run Figure 40 shows that the last pette is detected at position 10 and a wro...

Page 21: ...pettes are loaded from position 1 to 4 However the Operator entered position 6 for the pette start position refer to section 2 9 Figure 42 Alarm Screen of Communication Error Screen Display Figure 42 Summary After the power is turned on Figure 42 will appear if the control software cannot communicate with the panel operation The version information and current date time will not be displayed in th...

Page 22: ...Auto Run Pause Maintenance AP720S Panel Flowchart Interrupted run 21 ...