Tube Voltage: Tube voltage can be specified from 4kV to 30kV. Type in the
5.7.3.3.
desired voltage, then tap
to accept the input and return to Test Measurement menu.
Tube current: Tube current can be specified in 1uA steps from 10uA to 750uA.
5.7.3.4.
The maximum value may be limited by the tube voltage previously specified. Type in the
desired tube current and tap
to accept the input and return the Test Measurement
menu.
Filter: For those systems with filters, choose which, if any, filter should be used.
5.7.3.5.
Select a filter from the list of 5 possible filters, then tap
to accept the input and return
to the Test Measurement menu.
There are three filter options on the LAB-X5000, depending on options chosen
and which X-ray tube is installed.
1.
No filters.
2.
Ti: (0) Open; (1) W1 Ti; (2) A6 thick Al; (3) Z1 Pd; (4) KF Kapton
3.
Pd: (0) Open; (1) A4 Al ; (2) A6 thick Al; (3) Z1 Pd; (4) Z1 Pd KF Kapton