21
) Q1E-EP1370
8. Specifications
8.1
Probe
Type :
L44
Probe
Acoustic working
frequency :
6.5MHz
Technology
: Linear Array Probe
Dimensions
: See Fig. 9.
Weight :
Approx.
0.43kg
(Including cable and connector)
Scanning width
: 38mm
Probe materials
: Biocompatible allergy free components
Acoustic output
: According to IEC60601-2-37
(See Main Unit manual.)
Applicable system
: Depending on production and upgrade status
for detailed information, contact a service
support.
Classification
: MDD classification IIa.
Cleaning
: Applicable detergents are listed in the
suppliers list
Disinfection :
Applicable
disinfectants are listed in the
suppliers list
Sterilization
: ETO gas sterilization
Plasma
sterilization
Operating conditions:
Ambient temperature
: +10 - +35°C
Contact surface temperature : max. 42°C
(Temperature of examinee)
Relative humidity
: 30 - 85%
Storage conditions:
Temperature
: -10 - +55°C
Relative humidity
: 10 - 95%
(subject to no condensation)