) Q1E-EP1370
5. Reprocessing
Procedure
The probe must be reprocessed after each use. Refer to the
reprocessing instruction in this chapter. For reprocessing of the
c
oupler attachment EZU-PA7L1 and needle guide refer to the
instruction manual of EZU-PA7L1.
WARNINGS
‐
The probe is delivered unsterile. Prior to the first
use, reprocess the probe.
‐
Temperature should not exceed 60°C during
reprocessing.
‐
Probe connector is not water resistant.
Limitations on
reprocessing
The probe is not completely submersible. The
immersible part is shown in Fig.1. The un-immersible
part should be disinfected by wipe disinfection.
Transportation
before using
The probe should be packed in a sterile pouch or
container to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful not
to damage the sterile pouch or container during
transportation.