5.7.
Sterilization
The probe and accessory can be sterilized using either ethylen oxide
gas (EtO) sterilization or plasma sterilization (see table in the next
page). Follow the manufacturer's instructions of the sterilizer
regarding usage, temperature and sterilization-time.
The sterilization method and operating conditions are as follows.
Sterilization Method
Condition
Plasma Sterilization:
STERRAD® 50, 100S or 200 (*) Short Cycle
Plasma Sterilization:
Sterrad® NX or 100NX (*)
Standard cycle
ETO Sterilization
Gas Type: 10% EO/ 90% HCFC
Temperature: 50-55°C
Exposure Time: More than 120 minutes
Pressurization: 162-200kPa
Depressurization: 13-8kPa
Relative humidity: 40-90%
Aeration is minimum 12 hours
* STERRAD® systems are manufactured by "Johnson & Johnson"
1)
Before performing sterilization, check that the operation data of
sterilizer are in conjunction with min. and max. data applicable
for the probe and the Attachment.
2)
Do not sterilize the probe and the Attachment by Steam Autoclaving.
If you autoclave them, they suffer serious damage and will not be
functional.