-7-
Q1E-EP0325
6)
Fix jelly on the top of the probe with
sterile adhesive tape. (See Fig. 5.)
7)
Insert the probe gentry and adjust the probe’s position for a clear
view of the desired image. By turning or tilting the probe, a “turn
around” imaging is possible. If the image is unsufficient, use
sterile saline solution. This solution improves the contact between
probe and organ.
8)
After using the probe, clean, disinfect and if necessary sterilize
the probe immediately and using RVS, also clean, disinfect and
sterilize the Magnetic sensor attachment. In case of using RVS with
an EZU-RV2S, also clean, disinfect and sterilize the Spacer for
EZU-RV2S.
9)
Store the probe, the Magnetic sensor attachment and the Spacer for
EZU-RV2S in the environment indicated in “6. Maintenance and Safety
Inspection”.
3.2.
Use of Sterile Puncture Adapter (EZU-PA5V)
The process of attaching the Sterile Puncture Adapter (EZU-PA5V) to
the probe is as follows. If the Sterile Puncture Adapter is used, careful
handling is necessary to avoid damage of the probe cover. During open
or minimal invasive surgery, use the protection sleeve to cover probe
and cable.
1)
Attach a sterile probe cover to the probe. (See “3.1 Connection and
Settings”)
2)
Put the picks of the Sterile
Puncture Adapter to the
grooves on the tip of the
probe. (See Fig. 6.)
3)
Push the other end of the
Sterile Puncture Adapter until
fix the dents of the Sterile
Puncture Adapter to the
projection on the probe. (See
Fig. 7.)
Sterilized tape
Fig. 5
Sterile Puncture Adapter
(EZU-PA5V)
Probe cover
Fig. 6
Fig. 7