119
Test Time (EOM) Display
8.9 Test Time (EOM) Display
Displays the test time (EOM) for 1 test.
Test time= (INDEX + software processing time + judgment times) × Number of pulse applications
(Measurement screen)
[SYSTEM]
>
[SYSTEM]
1
1
Tap
[EOM TIME]
, and set whether to display or hide the test times (EOM).
OFF
Test times are not displayed.
ON
Test times are displayed.
2
Implement the test.
After the test finishes, the test time is displayed in the top right of the screen.
IMPORTANT
The test times displayed onscreen are rough estimate times calculated by the internal software, and there is
a risk that they will differ considerably from the external control terminal (EXT. I/O) EOM signal outputs.
Additionally, if the automatic save setting is enabled, the displayed time will not include the time required for
save processing. For accurate EOM times, check using external control terminal EOM signal outputs.