Set and query the smoothing function
Syntax
Command
[:SENSe]:AVERage: COUNt
<1 to 100(NR1)> [,<1 to 12(NR1)>]
Query
[:SENSe]:AVERage: COUNt?
[<1 to 12(NR1)>]
Response
<1 to 100(NR1)>
Description
Sets or queries the average number of times for the channel specified using the last
parameter.
Command: If channel numbers are omitted, the same settings are made in all the
channels.
Query: If parameters are omitted, the settings of all channels separated by commas are
returned.
Example
:AVER: COUN 10,1
Set the number of averaging iterations for channel 1 to 10.
:AVER: COUN 10
Set the number of averaging iterations for all channels to 10.
:AVER: COUN? 1
10
The setting is made so that channel 1 performs the averaging 10 times.
(9)
Logging
Start saving the voltage and current values
Syntax
Command
:DATA: STATe
<1/0/ON/OFF>[,<1.00 to 99.99 (NR2)>]
Query
:DATA: STATe?
Response
<1/0(NR1)>
Description
Starts saving the current and voltage values for all channels to the instrument memory.
Specifying the saving time (1.00 s to 99.99 s) will stop the saving after the specified time
has elapsed from the start of saving. When the saving time is omitted, the current and
voltage values are saved for 12 hours continuously until the operation stops.
In addition, an execution error occurs during saving.
Example
:DATA: STAT 1,5.00
Starts saving and stops it after 5 seconds.
:DATA: STAT?
1
Query the number of saved data
Syntax
Query
:DATA: POINts?
<1 to 12 (NR1)>
Response
<0 to 15000(NR1)>
Description
Returns the number of data that is currently saved by the channel specified using the
parameter.
Example
:DATA: POIN? 1
120
Query the saved voltage data
Syntax
Query
:DATA: VOLTage?
<1 to 12 (NR1)> [,<1 to 15000 (NR1)>]
Response
<Measured value (NR3)>, <Measured value (NR3)>••••<Measured value
(NR3)>
Description
Specifies the channel and the number of data, and reads the voltage values that have
been saved.
All data points are returned if the number of data is not specified.
If any of the following occurs, an execution error occurs.
• While saving
• If the number of read data points is greater than the number of data points saved
• If no saved data is found
Do not issue commands or queries while data is being transferred. To cancel a data
transfer, execute
*
CLS
.
Example
:DATA: VOLT? 1,10
+4.00,+4.00•••+4.00
HIOKI SS7081C961-00
Summary of Contents for SS7081-50
Page 7: ...4 HIOKI SS7081C961 00...
Page 13: ...10 HIOKI SS7081C961 00...
Page 21: ...18 HIOKI SS7081C961 00...
Page 57: ...54 HIOKI SS7081C961 00...
Page 65: ...62 HIOKI SS7081C961 00...
Page 67: ...64 HIOKI SS7081C961 00...
Page 69: ...66 Appx 2 Dimensions Diagram 480 460 430 483 15 57 1 Unit mm HIOKI SS7081C961 00...