45
Troubleshooting
5.1 Troubleshooting
If damage is suspected, contact your authorized Hioki distributor or reseller after reviewing the
contents below.
CAN signal detection is unstable.
Check/consider
Solutions
See
•Have the Signal Probes been connected properly
to the cable (CAN bus) under test?
•Verify that the cable (CAN bus) under test has
been properly inserted between the detection
electrode and guard hook.
•If the Signal Probe has not been connected
(tightened) properly, the product may not be able
to detect signals stably.
•Review the method for connecting the Signal
Probes.
•Is there any dirt on the Signal Probe’s detection
electrode or guard hook, or on the location on
the cable (CAN bus) under test where the Signal
Probe will be affixed?
•Remove any dirt from the Signal Probe’s
detection electrode and guard hook.
•Wipe away any dirt from the location on the cable
(CAN bus) under test where the Signal Probe will
be affixed.
–
•Has the ground connection cable been properly
connected?
•Connect the ground connection cable to the
CAN signal source ground or to the vehicle’s
body (ground connected to the vehicle battery’s
negative terminal or body ground).
•You will not be able to detect CAN signals
properly if the ground connection cable is
connected to a metal object that is insulated
from the vehicle’s ground (ground connected to
the vehicle battery's negative terminal or body
ground). If you are unsure where to connect the
ground connection cable, first check for continuity
between the metal part you are considering and
the vehicle’s body ground and then connect the
ground connection cable.
•Does the SP7100 CAN Interface’s
CH1
or
CH2
LED flash red/green alternately at high speed?
•Check the settings for the device to which the
CAN Interface’s output is connected to see if that
device may be outputting a signal.
•
If the LED flashes red/green alternately at high
speed, it indicates that the device to which the
CAN Interface’s output is connected may be
outputting a signal.
•Does the SP7150 CAN Interface’s
STATUS
LED
flash red/green alternately at high speed?
HIOKI SP7001A961-01
Summary of Contents for SP7001
Page 2: ...HIOKI SP7001A961 01 ...
Page 4: ...Contents ii HIOKI SP7001A961 01 ...
Page 30: ...26 Inspecting the Product Before Use HIOKI SP7001A961 01 ...
Page 34: ...30 Output Bus Error Detection Function HIOKI SP7001A961 01 ...
Page 51: ...HIOKI SP7001A961 01 ...
Page 52: ...HIOKI SP7001A961 01 ...