3.1 Pre-Operation Inspection
17
Setting Measurement
Conditions
Chapter 3
3.1 Pre-Operation Inspection
Do not use a damaged cord because
doing so may result in electric
shock. (Replace the cord with a new
one.)
Metal Exposed
•
Is the power cord insulation torn, or is
any metal exposed?
•
Is the connection cord insulation torn, or
is any metal exposed?
1
No Metal Exposed
Before using the instrument the first time, verify that it operates normally to ensure that the
no damage occurred during storage or shipping. If you find any damage, contact your dealer
or Hioki representative.
Peripheral Device Inspection
If damage is evident, request re-
pairs.
Yes
Is damage to the instrument evident?
When turning power on
Do the “HIOKI”, “SM7810” and version num-
ber indications appear on the display area?
No
2
The power cord may be damaged, or
the instrument may be damaged in-
ternally.
Request repairs.
No
Yes
Inspection complete
Please read the "Operating Precautions" (p.4) before use.
Use of a supply voltage outside the
specified range may damage the in-
strument or cause electric shock.
Before turning the instrument on,
make sure the supply voltage
matches that indicated on its power
connector.
No
Does the supply voltage of your power
source match the supply voltage indicated
above the power source inlet on the rear of
the instrument?
Yes
Instrument Inspection