Example: Measurement with Model 8966 Analog Unit

If the power voltage is measured with 20 V/div of the vertical axis range and ×1 of the vertical axis 

magnification settings, the minimum resolution is calculated as follows:

Screen full scale: 20 V/div × 20 div = 

400 V

Full scale resolution at ×1 of the vertical axis magnification: 

2000

Minimum resolution: 

400 V / 2000 = 0.2 V

Full-scale resolution for input modules at each vertical axis magnification (LSB)

Input module

Magnification

×1/10

×1/5

×1/2 

×1

×2

×5

×10

×20

×50  ×100

8966 

(Analog)

8971 

(Current)

8972 

(DC/RMS)

20000

(4000)

10000

(4000)

4000

2000 

1000

400

200

100

40

20

8967 

(Temperature)*

200000

100000

40000

20000

10000

4000

2000

1000

400

200

8968 

(High resolution)

U8974

 

(High voltage)

U8975

(4ch analog)

U8978

(4CH analog)

U8977

(3CH current)

320000

(64000)

160000

(64000)

64000

32000

16000

6400

3200

1600

640

320

8969, U8969 

(Strain)

U8979

(Charge)

250000

(64000)

125000

(64000)

50000

25000

12500

5000

2500

1250

500

250

8970 

(Power frequency) 20000

10000

4000

2000

1000

400

200

100

40

20

8970 

(Integration)

400000

200000

80000

40000

20000

8000

4000

2000

800

400

8970 

(Excluding power 

frequency and 

integration)

100000

50000

20000

10000

5000

2000

1000

500

200

100

MR8990 

(DVM)

1200000 1200000  1200000  1000000  500000  200000  100000 50000  20000  10000

The figures enclosed in parentheses indicate valid data range.

*: For Model 8967 Temp Unit, the valid ranges differ depending on types of thermocouple. For information on 

the minimum resolution, see the specifications of Model 8967 Temp Unit in the instruction manual.

Data Saving Speed

The table lists the speeds of saving binary data 

(values used as a reference) using different media 

and interfaces. Note that the data saving speed 

varies depending on the saving conditions, device 

manufacturer, device capacity, communication 

conditions, and others.

Storage media

Saving speed

(As a reference)

CF card

800 kB/s

Built-in drive (SSD)

800 kB/s

USB flash drive

800 kB/s

Computer via LAN

1 MB/s

HIOKI MR8847G963-03

Summary of Contents for MR8847-51

Page 1: ...troduction to the Memory HiCorder s basic measuring method for first time users MR8847A MR8847 51 MR8847 52 MR8847 53 Measurement Guide MEMORY HiCORDER EN Apr 2021 Revised edition 3 MR8847G963 03 21 0...

Page 2: ...HIOKI MR8847G963 03...

Page 3: ...orm Generator Unit MR8791 Pulse Generator Unit and SF8000 Waveform Maker Contents Screen structure and operation overview The screen structure and overview of the operation keys of the instrument p 2...

Page 4: ...memory internal memory Settings window Status screen Displayable on the waveform screen Channel settings window Trigger settings window Sheet tabs Displays selectable sheet names The screen to set mea...

Page 5: ...e running operation Closes the displayed dialog and the window KEY LOCK Press and hold ESC key for 3 s to disable key operations Press and hold it for 3 s enable key operations 12 WAVE key Lit in red...

Page 6: ...19 FEED key Feeds the recording paper 23 TIME DIV key Configures the time axis setting 20 SAVE key Lit in blue while accessing a recording medium Saves data a dialog can be switched between visible an...

Page 7: ...e the specified clearances around the instrument All around the instrument at least 5 cm Left side Vent holes Set the items necessary to the instrument 1 Connect the probes and cords 1 2 4 2 3 Refer t...

Page 8: ...fferences and sets the reference potential of the instrument to 0 V To ensure precise measurements perform zero adjustment after 30 minute warm up 1 2 Configure the setting on the Unit List sheet 1 Mo...

Page 9: ...leads or clamp sensor to the measurement point Set the measurement conditions Configure the setting on the waveform screen Measurement function Timebase horizontal axis Recording length On the Channe...

Page 10: ...yed waveforms can be changed It can be stretched or compressed Save Save the data The data can be also automatically saved under the pre configured saving conditions Observe the saved data The data ca...

Page 11: ...198 Connection Cord CF card Refer to Preparations for Measurement p 5 Insert the CF card Connect the lead to the measuring point Connect the power cord to the power outlet Connect the lead to CH1 2 Se...

Page 12: ...ed as follows Move the cursor Select the items to be set Displays the File Save sheet on the system screen Set the saving conditions Select at save Selection Save No Pressing the SAVE key allows savin...

Page 13: ...gth equivalent to the preset record length will be recorded Measurement data will be recorded until the STOP key is pressed 5 Save the data Press SAVE key Data will be saved under the pre set saving c...

Page 14: ...Refer to Preparations for Measurement p 5 Insert the CF card Connect the lead to the measuring point Connect the power cord to the power outlet Connect the lead to CH1 2 Set the measurement condition...

Page 15: ...sheet Auto Save Set to On Save to Specify the folder the waveform to be saved to For automatic saving the built in drive a USB flash memory and a CF card are available When saving waveforms to the CF...

Page 16: ...ayed on the display is indicated on the scroll bar Red bar Whole waveform Blue frame Currently displayed range Trigger point Trigger time Earlier Later Stretch or compress the waveform Change the magn...

Page 17: ...sured value 1 Set the A and B cursors Configure the setting on the waveform screen as follows Display the waveform screen Move the cursor Select the items to be set 1 2 Displays the AB cursor settings...

Page 18: ...cursor only a B Use the A and B cursors Move the B cursor only A B Use the A and B cursors Move the both simultaneously Display the waveforms eliminating overlap each other When multi phenomenon are m...

Page 19: ...ed or compressed in the vertical axis direction Zero position The waveforms can be placed at other positions in a vertical direction Can be changed even during measurement Positions of logic waveforms...

Page 20: ...he status screen Num Calc sheet Numerical Calc Set to On Parameter Set the calculation conditions Type Set the calculation type Channel Select the channel to be calculated Once measurement is complete...

Page 21: ...nnel 1 Displays the file screen 3 Select a medium 4 Up Down Select a file or folder Left Move to the superior folder Right Move to a subordinate folder 1 2 Select Change Media 3 Select a recording med...

Page 22: ...ata Numerical calculation result Text CSV S Text data Comment for printing Text TXT L Text data Waveform evaluation setting data Binary ARE S L Waveform evaluation areas settings data Waveform evaluat...

Page 23: ...to the printer such as the print density The setting does not have to be configured every time Move the cursor Select the items to be set Display the system screen Printer sheet Print Area Whole Wave...

Page 24: ...nge default A B Wave Print the waveform data in the range specified by the A and B cursors the A and B cursors will not be printed About Trig Print the waveform data corresponding to 10 divisions each...

Page 25: ...f Printing Title Upper limit value Lower limit value Channel Values at the A and B cursors A4 size printing The A4 size printing is possible with the same operation as for the report printing by chang...

Page 26: ...L9198 Connection Cord CF card Refer to Preparations for Measurement p 5 Insert the CF card Connect the lead to the measuring point Connect the power cord to the power outlet Connect the lead to CH1 2...

Page 27: ...e the part before the abnormal occurrence If the pre trigger is used waveforms including portions of those existing before the trigger point are recorded allowing you to detect signs before the abnorm...

Page 28: ...Probe or Model 9666 100 1 Probe Move the cursor Execute the probe compensation Displays the system screen Environment sheet 2 3 1 3 2 A 1 kHz 5 V square wave is output from the external output termina...

Page 29: ...which is 1 100 of the timebase two values are recorded the maximum and minimum values obtained during the specified sampling period With 10 ms div of the timebase setting one data point is 100 s When...

Page 30: ...64000 125000 64000 50000 25000 12500 5000 2500 1250 500 250 8970 Power frequency 20000 10000 4000 2000 1000 400 200 100 40 20 8970 Integration 400000 200000 80000 40000 20000 8000 4000 2000 800 400 89...

Page 31: ...HIOKI MR8847G963 03...

Page 32: ...HIOKI MR8847G963 03...

Reviews: