64
Select Input Channels
4. Vertical Axis
(Voltage Axis)
Zoom
Vertical axis (voltage axis) zoom-up or zoom-down settings can be selected
separately for each channel. These settings will be used for display and printout.
Zoom is based on the zero position. The measurement resolution does not change.
Refer to “7.5.3 Vertical Axis (Voltage Axis) Magnification” (p. 132).
Use the Variable function for user-specified zoom.
The waveform can be inverted by reversing plus/minus.
Refer to “8.6 Set the Waveform Position (Variable Function)” (p. 158).
Refer to “8.8 Invert the Waveform (Invert Function)” (p. 162).
5. Vernier
Fine adjustment of input voltage can be performed arbitrarily on the
Waveform screen (display only).
When recording physical values such as noise, temperature and acceleration
using sensors, amplitude can be adjusted to facilitate calibration.
Refer to “8.7 Fine Adjustment of Input Values (Vernier Function)” (p. 161).
6. Zero Position
Sets the 0 V level display position. If the 0 V input level has shifted, perform
zero-adjust.
Refer to “2.7 Adjust the Zero Position (Zero-Adjust)” (p. 47).
Refer to “2.8 Perform Calibration (When Model MR8990 is installed)” (p. 48).
The zero position of all channels can be changed at the same time using the
zero position preset (press
[Preset]
).
Ascending
(3%→96%)
The zero position of CH1 is set to 3% and the following channels
up to CH32 are increased by 3% each.
Descending
(96%→3%)
The zero position of CH1 is set to 96% and the following channels
up to CH32 are decreased by 3% each.
All Ch 0%
The zero position of all channels is set to 0%.
All Ch 50%
The zero position of all channels is set to 50%.
If the zero position of the waveform obtained with Model 8969 or U8969
Strain Unit is displaced, perform the auto-balance.
Refer to “8.10.4 Setting Model 8969 and U8969 Strain Unit” (p. 168).
• Only the display position is shifted. No offset is applied to the input.
• Vertical axis (voltage axis) zooming is based on the zero position.
• Zero position adjustment and vertical axis (voltage axis) zooming only change the voltage range as
displayed on the Waveform screen. The actual measurement range does not change.
The zero position is as shown in the illustration below. (Example: 8966 Analog Unit)
A/D Data
2047
A/D Data
0
0 V
2000 LSB
100 %
Display screen
(Zero position: 0%)
Display screen
(Zero position: 50%)
Display screen
(Zero position: 100%)
50 %
0 %
−2047
Zoom factor ×1
Full-scale resolution for modules at various vertical axis zoom factors (LSB)
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......