129
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
4
Alarm (Alarm Output)
This chapter describes how to set alarm conditions for each
measurement channel.
You can have the instrument sound a tone or output an alarm
signal to an external device when measurement data satisfies
the set condition.
For example, you can output an alarm when the recorded
temperature becomes too high.
Alarm signals can be outputted from the external control
terminals to external devices.
See “8 External Control (EXT. I/O)” (p. 173).
If an alarm condition is satisfied when measurement starts,
the alarm will be output immediately.
..............................................................
Setting shared alarm conditions for all channels ....................................
Configuring channel-specific alarm settings
...........................................
..................................................................
IMPORTANT
The logger is unable to detect alarm conditions when there is no data due to an interruption in
communication with wireless modules. However, it can detect the alarm condition of the wireless
module communication interruption.
When communication is reestablished with the wireless modules, the logger will restore data and
detect alarm conditions.
4 Alarm (Alarm Output)
4
Alarm (Alarm Output)
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...