Discharging Function
3.6 Discharging Function
After measurements are completed, discharge the measuring object.
When objects with capacitance component are measured, a charge
equivalent to the rated measurement voltage remains in the object
that may cause electric shock.
Even when the solar panel is discharged after measurement, a
generated voltage from the solar panel is detected that may not
be cleared.
OFF
Without removing the test leads
from the measurement object, turn
off the
MEASURE
key.
Discharging
Turned on
The built-in discharge circuit
automatically discharges the item.
In models IR4057 and IR4058, the bar
graph level decreases according to
discharge. However, measuring objects
with smaller capacitance component
discharge quickly and the bar graph level
may not change.
Discharged
Off
When the discharge is completed,
is turned off.
The time required for discharge depends
on the capacitance value.
1
2
4
5
6
7