Comparator Function
2
Set the rotary switch to the range for which you wish to set
the judgement reference.
Range
Operation
500 V
1000 V Press
to release the lock.
PV
Ω
Press
to select the applied voltage and press
to
release the lock.
Blinks
3
2
COMP
REF
or
3
Press .
[COMP]
or
[REF]
blinks and the
resistance value that will be used as
the judgement reference is displayed.
IR4053, IR4055, IR4056:
REF
IR4057-50:
COMP
4
4
Press or to select the
judgement reference.
If there is no operation for about 2
seconds after you select the judgment
reference, the comparator will be set,
and
[COMP]
or
[REF]
will light up.
Canceling the Comparator
oFF
Press several times to select
[oFF]
.
If there is no operation for for about 2
seconds after you select,
[COMP]
or
[REF]
will go off, and the comparator function will
be canceled.