20
―――――――――――――――――――――――――――――――――――――――
Chapter 5 3184-02 GP-IB Specifications
―――――――――――――――――――――――――――――――――――
Header
Function
V
12
V
12
(Voltage measurement)
V
32
V
32
(Voltage measurement)
A
1
A
1
(Current measurement)
A
3
A
3
(Current measurement)
W
W (Power measurement)
NO
Indicates no setting has been made.
OVR
Overrange
Measurement
function
Range
Mantissa
Exponent
V
12
, V
32
200 V
+ddd.d
E+0
500 V
+dddd.
E+0
A
1
, A
3
2 A
+d.ddd
E+0
20 A
+dd.dd
E+0
W
0.2 kW
±
ddd.d
E+0 *
2 kW
±
d.ddd
E+3
20 kW
±
dd.dd
E+3
Note: Data marked by asterisk (*) is formatted different
than the examples shown on the instrument’s
instruction label.
Output data for an overrange is as follows.
OVR 19999. E
+
9
5.4 Talker Function
Output data format
AAA
±
DDDDDE
±
D CR LF
(1)
(2)
(3)
(1) Measurement Functions
(2) Measurement Values
Summary of Contents for HITESTER 3184
Page 2: ......
Page 4: ......
Page 12: ...2 Chapter 1 Overview Front Panel Rear Panel 1 2 Names and Functions of Parts...
Page 16: ...6 Chapter 1 Overview S L I D E T O R O T A T E Handle...
Page 38: ...28 Chapter 5 3184 02 GP IB Specifications...
Page 41: ......
Page 42: ......