DT4281,DT4282 DIGITAL MULTIMETER    Remote Operation Manual 

P3/14 

1. Remote Interface Overview 

 
Step1.    Connect the USB cable to the USB port on the computer. 
Step2.    Once the USB cable is connected and the USB driver is installed, a virtual COM port 

is created on the computer. 

 
 
 

2. Interface Parameters 

 
In  order  to  operate  the  meter  via  a  host  computer  or  terminal,  the  parameters  in  interface 
within the DT4281,DT4282 have to match the parameters the serial interface provided by the 
host or terminal. 
The  following  procedures  will  guide  the  user  to  set  up  interface  parameters  within  the 
DT4281,DT4282 to comply interface with the host. 
 

Table.1 

Item 

Parameter 

Setting 

Baud Rate 

19200 bps 

Data Length 

8 bit 

Parity Check 

None 

Stop Bit 

1 bit 

 
 
 
 

Summary of Contents for DT4281

Page 1: ...DT4280 Series DIGITAL MULTIMETER Remote Operation Manual HIOKI E E CORPORATION Aug 2021 Revised edition 4...

Page 2: ...lt 4 3 3 Data Types 4 4 Summary of Commands 5 5 Detail of Commands 6 5 1 Description of Commands 6 5 2 Combination response string function and range 11 5 3 Index of threshold of continuity check 11 5...

Page 3: ...2 Interface Parameters In order to operate the meter via a host computer or terminal the parameters in interface within the DT4281 DT4282 have to match the parameters the serial interface provided by...

Page 4: ...tes a query command the return of the result will be in the following format Result CR LF 3 3 Data Types Returned message is the ASCII string from the meter responding to a query Table 2 Data Type Exp...

Page 5: ...value of the relative main CALC REL OFFS2 Query the offset value of the relative sub SYST APS 0 1 Set the APS SYST BEEP 0 1 Set the action of the beeper SYST BLIT 0 1 Set the backlight SYST BATT Query...

Page 6: ...ple HIOKI DT4281 121107517 Ver 1 00 SYST RST Put the meter to reset Syntax SYST RST Response OK SYST LLO Put the meter into the local lockout state Syntax SYST LLO Response OK SYST GTL Put the meter i...

Page 7: ...0 CALC REL OFFS CALC REL OFFS2 Query the offset value of the relative main or sub Syntax CALC REL OFFS CALC REL OFFS2 Response data 2 data offset value range Example 20 600m SYST APS Set the APS Synta...

Page 8: ...TAT Response ABCDEFGHIJKLMNOPQRSTUVWX 24char A Recording 0 OFF or 1 MAX or 2 MIN B Relative value REL 0 OFF or 1 ON C Filter 0 OFF or 1 ON D Beep 0 OFF or 1 ON E APS 0 OFF or 1 ON F Battery level 0 3...

Page 9: ...or 1 on Example SYST FILTER 1 Response OK or CMD ERR SYST PEAK Set the peak Syntax SYST PEAK data 1 data 0 off or 1 on Example SYST PEAK 1 Response OK or CMD ERR SYST SLOW Set the average Syntax SYST...

Page 10: ...e SYST CONDUCT 1 Response OK or CMD ERR SYST DIODE Set the threshold of diode test Syntax SYST DIODE data 1 data Index of threshold 0 6 Example SYST DIODE 1 Response OK or CMD ERR SYST DBM Set the sta...

Page 11: ...60 ACDCV 6 60 600 1000 SEPV 60m 600m 6 60 600 1000 CONT 600 DIODE 4 RES 60 600 6k 60k 600k 6M 60M 600M TEMP 800 CAP 1n 10n 100n 1u 10u 100u 1m 10m 100m CLAMP 10 20 50 100 200 500 1000 nS 600 DCuA 600...

Page 12: ...est Table 7 Index Threshold 0 0 15V 1 0 5V 2 1 0V 3 1 5V 4 2 0V 5 2 5V 6 3 0V 5 5 Index of standard impedance of dBm Table 8 Index Standard Impedance Index Standard Impedance 0 4 10 150 1 8 11 200 2 1...

Page 13: ...e of abnormal data Response of count value query return the values shown in Table 9 in the case of abnormal data Table 9 Type of abnormal value Count value Over Range value 1000000 Invalid data 200000...

Page 14: ...nto the local lockout state when in remote control This means no local key operation at the front panel is allowed during remote control GTL Put the meter into the local state clearing the remote stat...