16
_____________________________________________________________________
3.3 Setting by COMMUNICATION BASE
______________________________________________________________
NOTE
Previously recorded logger data is erased when recording is
resumed. Be sure to load data to be saved to
COMMUNICATION BASE or to personal computer before
recording.
At any other time even when interval setting display is not
shown, except during recording and waiting for recording,
communication with COMMUNICATION BASE is available.
However communication is disabled when logger is set to sleep.
3.3 Setting by COMMUNICATION BASE
1. Press logger INTERVAL button lightly to display LCD.
2. When logger LCD shows "
REC
" mark or clock icon,
press REC/STOP button for more than 1 second to stop
recording.
During recording or waiting time before recording start
time, data transfer cannot be established with
COMMUNICATION BASE.
3. Press logger INTERVAL button to display interval
setting display. ("
INTVL
" appears.)
4. Connect COMMUNICATION BASE with logger.
5. Press COMMUNICATION BASE SEND button for more
than 1 second to send data settings to logger.