![Hill-Rom The Vest P205 User Manual Download Page 73](http://html1.mh-extra.com/html/hill-rom/the-vest-p205/the-vest-p205_user-manual_2134210073.webp)
The Vest
®
Airway Clearance System, Model 205
67
User Manual (140643 REV 7)
PROPRIETARY AND CONFIDENTIAL DRAFT 19-DEC-11
Guidance and Manufacturer’s Declaration—
Electromagnetic Immunity
The Vest
®
Airway Clearance System, Model 205 is intended for use in the electromagnetic
environment specified below. The customer or user of the Model 205 should make sure it is
used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment—Guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Model 205, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 V
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2,5 GHz
3 V/m
80 MHz to
800 MHz
800 MHz to
2,5 GHz
where
P
is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and
d
is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
found by an electromagnetic site survey
a
,
should be less than the compliance level in
each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with this symbol:
d
3.5
3
-------
P
=
d
3.5
3
-------
P
=
d
7
3
---
P
=