![Hill-Rom The Vest P205 User Manual Download Page 31](http://html1.mh-extra.com/html/hill-rom/the-vest-p205/the-vest-p205_user-manual_2134210031.webp)
The Vest
®
Airway Clearance System, Model 205
25
User Manual (140643 REV 7)
PROPRIETARY AND CONFIDENTIAL DRAFT 19-DEC-11
2. Press the two
outside
Down
arrow buttons
for a minimum of 3
seconds (or until the
screen changes).
3. Press the
Down
arrow
button below
Spanish
(it will turn Spanish to
SPANISH).
4. To accept the change,
press the
Down
arrow
button below
ACEPTAR
(ACCEPT in English).
5. Once you have
accepted the change,
the system will automatically re-boot and go back to the main
screen in Spanish.
6. Once the main screen is visible, unplug the unit and plug it back in.
The unit will automatically reconfigure in Spanish.
7. To switch back to English, repeat the steps and choose
ingles
instead of
Spanish
.
Disable the Program Modes (Program and Ramp)
1. Press the left
Up
arrow button above
NORMAL
. The Normal
mode screen shows.
2. Press the
two outside
Down
arrow buttons
for a minimum of
3
seconds
, or until the
screen changes.
3. Press the right
Up
arrow button above
Programs
.