The Vest
®
Airway Clearance System, Model 205
41
User Manual (140643 REV 7)
PROPRIETARY AND CONFIDENTIAL DRAFT 19-DEC-11
4. Look at the settings on
the screen.
Ramp
Time Percent
is the
portion of the
treatment Time
required to increase
from the initial
settings (
Ramp Start
) to the final settings (
Ramp End
). To adjust
the percent to the prescribed therapy, press the right
Up
or
Down
arrow buttons.
5. Press the
ON
button to advance to the next screen.
6. Review the settings for Frequency and Pressure for the initial
setting,
Ramp Start
.
a.
To adjust the
Frequency
setting, press the
left
Up
or
Down
arrow buttons
until the
prescribed
treatment shows
on the screen. The Frequency (or Hertz (Hz)) may be set
between 5 and 20 cycles per second.
b. To adjust the
Pressure setting,
press the middle
Up
or
Down
arrow buttons
until the
prescribed
treatment
Pressure shows on the screen. Pressure may be set between 1
and 10.
7. Press the
ON
button to advance to the next screen.
8. Review the settings on the screen. Frequency and Pressure for the
final setting,
Ramp End
, are shown as treatment Time.
a.
To adjust the Frequency setting, press the left
Up
or
Down
arrow buttons until the prescribed setting shows on the screen.
Frequency or Hertz (Hz) may be set between the Ramp Start
setting and 20 cycles per second.