
84
Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
Troubleshooting
Caution Messages
Caution messages show on-screen to
let you know that an issue has
occurred. Follow the on-screen
instructions to troubleshoot the
issue. If an error code shows, see
“Solutions for Cautious Messages” on
page 84.
Solutions for Cautious Messages
Error Code
Do this:
007, 013, 020
1.
Press
return
after you read the on-screen
message.
2.
Power off the control unit, and then power
it on.
011
1.
Make sure the power cord is fully plugged
into the electrical inlet on the back of the
control unit. If necessary, disconnect the
power cord from the control unit, and then
connect it.
2.
Make sure the power cord is securely
plugged into an AC power.
3.
Press
return
after you read the on-screen
message.
012
1.
Make sure the foot switch buttons are not
pressed down or stuck.
2.
Press
return
after you read the on-screen
message.
014
1.
Make sure the correct patient circuit is set
up for the required therapy. For details, see
“Assemble and Connect the Patient Circuit”
on page 16.
2.
Press
return
after you read the on-screen
message.
Draft_E_2019-Apr-29_Cleaned
Summary of Contents for Synclara PSC1
Page 2: ...Draft_E_2019 Apr 29_Cleaned...
Page 109: ...Draft_E_2019 Apr 29_Cleaned...
Page 110: ...Draft_E_2019 Apr 29_Cleaned...