Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
Device Settings
NOTE:
The pressure management settings show:
•
Pressure Ceiling (PC)
—the maximum therapy pressure that can
be selected by the user in manual mode or when modifying a
preset plan.
•
Highest Programed Pressure (HPP)
—the highest therapy
pressure that has been programed in both automatic and
manual modes in this system.
2.
Press the
+
or
-
controls to select the required
PC
setting.
NOTES:
•
Each press of the control increases/decreases the
PC
setting by
five (5) units.
•
The PC ranges between 5 - 70 cmH2O and is limited to the HPP
gathered from the system.
•
The PC must be equal or greater than the HPP.
•
For example,
If HPP equals 45 cmH2O, then PC must be equal to or
greater than 45 cmH2O. The user can not select a PC setting
of 44 cmH2O or lower.
3.
When complete, slide the
Clinical
Access
setting to
OFF
. This prevents
unauthorized persons from adjusting the
settings.