4
7EN125103 Rev. 13 • 2020
www.hillrom.com
Safety Instructions
No modification of the product is allowed.
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the lift, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Use of the product adjecent to other equipment should be avoided because it could result in improper operation, if
such use is neccessary, observe and verify that the other equipment is operating normally.
Electromagnetic disturbance, may affect the lifting performance of the product. Modification using other parts than original
spare parts (cables etc.) may affect the electromagnetic compatibility of the product. Particular care must be taken when
using powerful sources of electromagnetic disturbance, such as diathermy, etc, so that, for example, diathermy cables are not
positioned on or near the lift.
If you have any questions, please consult the responsible assistive device technician or the supplier.
The lift is not suitable for use in the presence of flammable mixtures, for example, in areas where flammable goods are
stored.
Before lifting, always make sure that:
• the lift strap is not twisted or worn and can move in and out of the lift freely
• the lifting accessories are not damaged
• the lifting accessory is correctly and safely applied to the patient in order to prevent injuries
• the lifting accessories are properly attached to the lift
• the lifting accessories hang vertically and can move freely
• the sling bar latches are intact; missing or damaged latches must always be replaced
• the sling’s strap loops are correctly connected to the sling bar hooks when the sling straps are extended, but before the
patient is lifted from the underlying surface.
Incorrect attachment of sling to slingbar may cause severe injury to the patient.
Only use the Multirall
TM
overhead lift with a carriage, adapter, sling bar and other accessories approved by Liko
TM
.
Never leave a patient unattended during a lifting situation!
Before use, make sure that:
• the lift is assembled in accordance with the assembly instructions
• the lifting accessories are properly attached to the lift
• the batteries have been charged for at least 8 hours
• you have read the instruction for use for the lift and lifting accessories
• personnel using the lift are informed of the correct use of the lift and lifting accessories
• the lifting accessory is selected appropriately, in terms of type, size, material and design with regard to the patient’s needs.
Installation of carriages for Multirall
TM
overhead lift shall be made by personnel authorized by Liko
TM
in accordance with
the installation instructions and recommendations for the current lift system.
Intended use
This product is not intended to be used by the patient alone. Lifting and transferring a patient shall always be performed
with the assistance of at least one caregiver. This product is used as a means to perform the lift but is not in contact with the
patient; therefore we do not go into the various patient conditions in this instruction for use.
Contact your Hill-Rom representative for support and advice.
Multirall™ 200 overhead lift has been tested by an accredited testing institute