17
7EN125103 Rev. 13 • 2020
www.hillrom.com
In order to ensure maximum battery life, it is important to charge the
batteries regularly. We recommend that you charge the batteries after use
or every night.
The batteries are fully charged after approximately 8 hours. When fully
charged, batteries in good condition are sufficient for approximately 60 lift
cycles.
The connector cable of the battery charger cannot be repaired. If the
cable is damaged, it should be replaced by a new one!
Charging
1. Check to ensure that the emergency stops are not pressed in during
charging.
2. Place the hand control in its intended place on the charger.
3. Connect the charger to an electrical outlet (100-240 V AC).
The charging starts automatically. The green LED indicates that the
charger is connected to a power supply. The yellow LED indicates that
the battery is charging. When the battery is fully charged, the charger is
switched off automatically and the yellow LED turns off.
NOTE!
If the lift is not to be used for a longer period of time, the hand
control should be placed in the charger. If the charger is not connected to
a power supply, the emergency stop button should be pressed in order to
prevent the battery from discharging.
Charging the Batteries
1.
3.
2.
Green
Yellow
White
(no function)
Prod. No. 3126101- 3126104
The AC connector of the charger unit
must be easy accessible to the operator.
Please read the instruction for use before use.
Battery is charging when yellow led is lit.