Line Managers
Each head siderail has an integrated Line Manager on top of the siderail that
helps keep lines such as IV lines and suction lines together and away from the
articulating frame.
CPR Release Handle
The CPR release handles are located on both sides of the upper frame,
between the head and intermediate siderails. The CPR control quickly and
safely lowers the head-of-bed to a flat position, even when unplugged.
To activate, pull and hold until the head section is lowered to a flat position.
If the legs are elevated, continue to hold until you hear a single beep
confirming the bed is flat.
NOTE: If equipped with a Centrella™ MAX powered air surface:
• It will go into Max Inflate for 60-minutes.
• An alert will sound before 60-minutes has elapsed alerting the caregiver to
reset the timer or the surface will return to Normal mode.
• A countdown timer will show on touchscreen.
Patient Restraint Holders
There are four restraint holders on each side of the bed to secure a variety
of restraints such as a vest, lap belt, wrist or ankle restraints. We do not
recommend using these holders for Foley or other drainage bags due to the
bed’s low height, which allows for potential risk of bags touching the floor.
WARNING: Restraints must be attached to the articulating section of the bed
at the correct attachment points.