Preventative Maintenance
112
www.liko.com
7.0
Instructions regarding the check points for
Golvo 7000/7007
1. Drive the lift down and when the mast reaches the
bottom the current should not be more than 2 Amps.
2. Check the limit switch. Remove the plastic cover.
3. Remove the aluminium cover.
4. Check the Limiting switch by driving the lift down and
pressing the switch.
14 Limiting switch
15 Actuator
Replace the actuator with Spare Part No. 20090014 after 10 years
Actuator needs to be replaced after 10 years