DTS-60A O
PERATORS
M
ANUAL
54
5.2.1 Select a Test
The Select a Test menu displays the list of all the existed tests. These
are the Simple Test, the Standards and the User Defined Tests, Figure
4-3. An arrow at the right side of a test's name indicates that this is the
last selected test. The latter is the test that will be started after turning-
on the equipment.
Figure 4-3 –
Select Program Menu
SELECT TEST
IEC 156/95 <--
VDE 0370
IEC 156/63
ASTM D1816/84A
ASTM D877/87
TEST
CREATE
SELECT
EDIT
DELETE
STEP 1
From the TEST menu point to SELECT.
Then press the ENTER key.
STEP 2
The system displays all the available tests. The
default test is indicated with an arrow at the left
side.
Use the UP and DOWN keys to select a test for
execution.
UP
DOWN
SELECT TEST
IEC 156/95 <--
VDE 0370
IEC 156/63
ASTM D1816/84A
STEP 3
Press the ENTER key to execute the selected test.
From now on this test will be the default test.
ENTER
12:10 01/08/05 65 °F
12:10 01/08/05 65 °F
12:10 01/08/05 65 °F
ASTM D877/87
ENTER
Summary of Contents for DTS-60A
Page 1: ...DTS 60A Dielectric Oil Test Set Operator s Manual Version 1 20 HighVoltage INC ...
Page 3: ...DTS 60A OPERATORS MANUAL ...
Page 5: ...DTS 60A OPERATORS MANUAL ...
Page 7: ...DTS 60A OPERATORS MANUAL ...
Page 10: ......
Page 48: ...DTS 60A OPERATORS MANUAL 38 ...
Page 56: ...DTS 60A OPERATORS MANUAL 46 Figure 4 11 Test Program Execution Example ...
Page 57: ...DTS 60A OPERATORS MANUAL 47 ...
Page 81: ...DTS 60A OPERATORS MANUAL 71 ...
Page 87: ...DTS 60A OPERATORS MANUAL 77 ...
Page 111: ...DTS 60A OPERATORS MANUAL 101 ...