DTS-60A O
PERATORS
M
ANUAL
16
1.6 Startup the Equipment
1.
Ensure that setup has been completed as noted in
paragraph Setup the equipment (1.4).
2.
Turn on Main Power.
3.
The equipment initiates a self-test, Figure 1.13. In case of
malfunction, an error message (see paragraph 1.3.4) is
displayed and the system will be halted.
...SELF TEST 1
12:10 01/08/05 65 °F
MENU
BACK
DOWN
UP
ENTER
INC
TM
HIGH VOLTAGE
Figure 1.13 – Startup and Self Test
4.
After the self-tests are completed, the unit proceeds with
the last selected test (Simple Test, Standard or User-
Defined).
5.
The unit will display the Electrode Warning message (see
paragraph 1.3.3) and it will wait for ‘ENTER’ to be pressed.
6.
The user may open the lid, check the electrodes (type and
spacing) and finally fill the appropriate test vessel with
insulating oil.
7.
After closing the lid, the user should press the ‘ENTER’ key
in order to initiate the test procedure.
Whenever you see an ERROR message
contact with the service or the
distributor.
Summary of Contents for DTS-60A
Page 1: ...DTS 60A Dielectric Oil Test Set Operator s Manual Version 1 20 HighVoltage INC ...
Page 3: ...DTS 60A OPERATORS MANUAL ...
Page 5: ...DTS 60A OPERATORS MANUAL ...
Page 7: ...DTS 60A OPERATORS MANUAL ...
Page 10: ......
Page 48: ...DTS 60A OPERATORS MANUAL 38 ...
Page 56: ...DTS 60A OPERATORS MANUAL 46 Figure 4 11 Test Program Execution Example ...
Page 57: ...DTS 60A OPERATORS MANUAL 47 ...
Page 81: ...DTS 60A OPERATORS MANUAL 71 ...
Page 87: ...DTS 60A OPERATORS MANUAL 77 ...
Page 111: ...DTS 60A OPERATORS MANUAL 101 ...