50
After a breakdown occurs, the system will idle for this programmable
duration.
f.
After Breakdown Stirrer Timer (sec) -ABDST
After a breakdown occurs, the system will enable the stirrer for this
programmable duration.
g.
After Breakdown Stage (0…29, next, end) - ABDSTG
After the above timers the test will proceed to another stage. This
programmable parameter indicates the next stage. The value can be
any following stage (eg. If current stage is 10 then the value could be
11 to 29), next if it is desired to continue to the next stage, or end if
it is desired to stop the test.
h.
Store Breakdown Voltage (yes/no)
If the user selects the yes option, the detected breakdown voltage will
be calculated to the statistics. Otherwise, the detected breakdown
voltage will NOT be calculated into the statistics.
In the following paragraph there is a detailed description of the Test
Program Operation, thus the reader will understand the exact meaning of
the above parameters.
Summary of Contents for DTS-60A
Page 1: ...DTS 60A Dielectric Oil Test Set Operator s Manual Version 1 20 HighVoltage INC ...
Page 3: ...DTS 60A OPERATORS MANUAL ...
Page 5: ...DTS 60A OPERATORS MANUAL ...
Page 7: ...DTS 60A OPERATORS MANUAL ...
Page 10: ......
Page 48: ...DTS 60A OPERATORS MANUAL 38 ...
Page 56: ...DTS 60A OPERATORS MANUAL 46 Figure 4 11 Test Program Execution Example ...
Page 57: ...DTS 60A OPERATORS MANUAL 47 ...
Page 81: ...DTS 60A OPERATORS MANUAL 71 ...
Page 87: ...DTS 60A OPERATORS MANUAL 77 ...
Page 111: ...DTS 60A OPERATORS MANUAL 101 ...