INOTEST
6
The supplementary rails
17
enable the measuring
range of the instrument to cover smaller sizes than
is otherwise possible: 35 – 280 mm interior, 15 –
255 mm exterior.
Assemble as follows, referring to Fig. 6:
2.4
Supplementary base
rails
– Fix the pair of base rails
17
to the measuring
head
1
and lock with fixing screws
18
(Fig. 7).
– Fit the adjustable fixed anvil
19
. Clamp one of the
measuring arms
4
into position on the mobile
anvil
3
.
– Insert the dial gauge as described in para. 2.1.
As the INOTEST is a comparative measuring in-
strument, it must first be set for the size in question
with the aid of a master gauge. The instrument only
indicates the deviation for set size.
Master gauges should be made of a material having
approximately the same coefficient of expansion as
the workpiece material.
Right
Wrong
3 ZERO SETTING
It is important to calibrate the instrument in the
same position as it will occupy when measuring the
workpiece.
3.1
Basic instrument
Fig. 6.
Fig. 7.
18
18
17
19
4
3