13.4.5  Brake de-energisation speed 

• 

Press the 

 key repeatedly until the parameter n

(•)

/RPM is displayed. The input field will be displayed inversely. 

• 

Set the desired value by means of the control knob   . 

 
 

13.5 Radius 

• 

Press the   key repeatedly until the parameter r/mm is displayed and the input field is shown inversely. 

• 

Set the desired value by means of the control knob   . 

The value of the RCF/RZB adapts automatically when the radius is changed; this is shown by the indication flashing.  
 
 

13.6 Automatic 

buffer 

The buffer comprises the program locations "

----

" and 90 to 99. Every time a centrifugation run is started, modified 

centrifugation data are automatically stored in program location "

----

". The modified centrifugation data for the last 11 

centrifugation runs are stored in the buffer and can be called up (see Chapter "Calling up programs"). 
 
 

14 Programming 

 

14.1  Entering / modifying programs 

• 

Set the desired parameters (see Chapter "Entering centrifugation parameters"). 

• 

Press the 

PROG

 key in order to select the parameter PROG-No. The input field will be displayed inversely. 

• 

Set the desired program location by means of the control knob   . 
If the program location indicator flashes, then this program location has already had centrifugation data assigned 
to it. In this case, set a free program location or overwrite the centrifugation data by continuing. 

• 

Press the 

STO

 key once to save the settings in the desired program location. 

Press the 

STO

 key twice to overwrite centrifugation data already stored. 

 
 

14.2  Calling up programs 

• 

Press the 

PROG

 key in order to select the parameter PROG-No. The input field will be displayed inversely. 

• 

Set the desired program location by means of the control knob   . 

• 

Press the 

RCL

 key. The centrifugation data for the selected program location will be displayed. 

 
 

15 Centrifugation 

 

If the permissible weight difference within the rotor loading has been exceeded, the drive shuts down during 
the start-up, and 

IMBALANCE / UNWUCHT

 is displayed. 

If the speed in the selected program is higher than the maximum speed of the rotor (n-max-Rotor), it will not 
be possible to start a centrifugation run. The message 

N > ROTOR MAX 96

 will be displayed (see Chapter 

"Faults"). 
A centrifugation run can be stopped at any time by pushing the key 

STOP

 .  

It is possible to switch over between RPM and RCF display at any time using the 

n

 and 

RCF

 keys. When 

working with the RCF display, it is necessary to enter the centrifuging radius. 
If 

 OPEN   OEFFNEN

 is displayed, a further operation of the centrifuge is only possible after opening the lid 

once. 
If 

R xx n-max xxxxx

 is displayed, then no centrifugation run has taken place as the rotor was changed 

beforehand, refer to Chapter " Rotor Identification ". 

• 

Switch on the mains supply switch. Switch position 

Ι

• 

Load the rotor and close the centrifuge lid. 

 
 

15.1 Centrifugation 

with time selection 

• 

Set the time or call up a program with a pre-set time (see Chapter "Entering centrifugation parameters" or 
"Calling up programs"). 

• 

Press the 

START

 key. The rotation indicator   will appear as long as the rotor is turning. 

• 

When the time has expired or if the centrifugation run is interrupted by pressing the 

STOP

 key, run-down will take 

place with the selected run-down parameters. When the rotor comes to a stop, the 

 symbol will flash in the 

display until the lid is opened.  

During the centrifugation run, the speed of the rotor or the RCF value resulting from this and the remaining time will 
be displayed.  
 
 

Summary of Contents for ROTANTA 460 S

Page 1: ...Rev 01 03 10 Andreas Hettich GmbH Co KG AB5601 02EN ROTANTA 460 S EN Operating Instructions 5 ...

Page 2: ...2 35 Fig 1 LOCK1 STOP 120 30 4000 200 22 PC Fig 2 ...

Page 3: ...ri è conforme alle direttive CE e alle norme citate Geräteart Type of device Type d appareil Tipo di apparecchio Laborzentrifuge mit Zubehör Laboratory centrifuge with accessories Centrifugeuse de laboratoire avec des accessoires Centrifuga da laboratorio con accessori Typenbezeichnung Type designation Désignation de modèle Contrassegno tipo ROTANTA 460 S EG Richtlinien Normen EC guidelines standa...

Page 4: ... 0 Fax 49 0 7461 705 125 info hettichlab com service hettichlab com www hettichlab com 2005 by Andreas Hettich GmbH Co KG All rights reserved No part of this publication may be reproduced without the prior written permission of the copyright owner Modifications reserved AB5601 02EN Rev 01 03 10 ...

Page 5: ...ing time 14 13 1 1 Time selection 14 13 1 2 Continuous running 14 13 2 Speed RPM 14 13 3 Relative centrifugal acceleration RCF RZB 14 13 4 Run up and run down parameters 14 13 4 1 Run up step 14 13 4 2 Run up time 14 13 4 3 Braking step 14 13 4 4 Run down time 14 13 4 5 Brake de energisation speed 15 13 5 Radius 15 13 6 Automatic buffer 15 14 Programming 15 14 1 Entering modifying programs 15 14 2...

Page 6: ...f materials or mixtures of materials with a density higher than 1 2 kg dm 3 19 30 Rotor Identification 19 31 Emergency release 20 32 Maintenance and servicing 20 32 1 Centrifuge housing lid and centrifuging chamber 20 32 1 1 Surface cleaning and care 20 32 1 2 Surface disinfection 20 32 1 3 Removal of radioactive contaminants 20 32 2 Rotors and Attachments 21 32 2 1 Cleaning and care 21 32 2 2 Dis...

Page 7: ... third parties and impairments to the machine or to other material assets The machine is only to be used for the specified applications and only in an impeccable technical safety condition Disturbances that can interfere with the safety are to be immediately rectified 3 Technical specifications Manufacturer Andreas Hettich GmbH Co KG D 78532 Tuttlingen Model ROTANTA 460 S Type 5601 02 Mains voltag...

Page 8: ...not be moved or knocked during operation In case of fault or emergency release never touch the rotor before it has stopped turning To avoid damage due to condensate when changing from a cold to a warm room the centrifuge must either heat up for at least 3 hours in the warm room before being connected to the mains or run hot for 30 minutes in the cold room Only the rotors and accessories approved b...

Page 9: ...to do so by the manufacturer Only original spare parts and original accessories licensed by the Andreas Hettich GmbH Co KG company are allowed to be utilised The following safety regulations apply EN IEC 61010 1 and EN IEC 61010 2 020 as well as their national deviations The safe operation and reliability of the centrifuge can only be guaranteed if the centrifuge is operated in accordance with the...

Page 10: ...e machine and in this document Symbol for the separate collection of electric and electronic devices according to the guideline 2002 96 EG WEEE The device belongs to Group 8 medical devices Applies in the countries of the European Union as well as in Norway and Switzerland 6 Delivery checklist 1 Connecting cable 1 Hex pin driver 2 5 mm 1 Hex pin driver 5 mm 1 Lubricating grease for trunnions 1 Ope...

Page 11: ...d Remove the transport safety device see instruction sheet on Moving the equipment safely 9 Opening and closing the lid 9 1 Opening the lid The lid can only be opened when the centrifuge is switched on and the rotor is at rest If it cannot be opened under these circumstances see the section on Emergency release Press the button on the front panel The lid unlocks via the motor the illuminated pushb...

Page 12: ... loading includes the total weight of adapter frame centrifuging container and content In containers with rubber inserts the same number of rubber inserts must always be among the centrifuge containers Always fill the centrifuge containers outside of the centrifuge No liquid should be allowed to enter the centrifugal chamber during filling and swinging out of the hangers The maximum filling quanti...

Page 13: ...st run up time Step 1 longest run up time 2 Run up time parameter min sec t The time range that can be set is dependent upon the set speed Run down parameters 1 Braking steps parameter 0 9 R linear braking curve B similar to an exponential braking curve Step R9 B9 short run down time Step R1 B1 long run down time Step R0 unbraked run down 2 Run down time parameter min sec t The time range that can...

Page 14: ...stment is being made 13 4 Run up and run down parameters The run up and run down parameters that have been set are displayed x 1 9 run up step t run up time x y PROFIL y R1 R9 B1 B9 braking step R0 unbraked run down t run down time n brake de energisation speed 13 4 1 Run up step Press the key repeatedly until the parameter 1 9 or t is displayed The input field will be displayed inversely Set the ...

Page 15: ... inversely Set the desired program location by means of the control knob Press the RCL key The centrifugation data for the selected program location will be displayed 15 Centrifugation If the permissible weight difference within the rotor loading has been exceeded the drive shuts down during the start up and IMBALANCE UNWUCHT is displayed If the speed in the selected program is higher than the max...

Page 16: ...playing the maximum speed of the rotor Press the n key repeatedly until the parameter RPM is displayed and the input field is shown inversely Press the n key once more and hold it pressed the maximum speed of the rotor n max Rotor will be displayed 19 Displaying the maximum RCF of the rotor Press the RCF key repeatedly until the parameter RCF RCA is displayed and the input field is shown inversely...

Page 17: ...s 24 Centrifugation data displayed after switching on After switching on the centrifugation data for program 1 or for the program last used will be displayed This can be set as follows Open the lid Switch the mains supply switch off and on again Switch position Ι Press the STOP key when the first visual change appears in the display inverse display PROGRAM 1 LAST PROGRAM will be displayed Set the ...

Page 18: ...desired parameters see Chapter Entering centrifugation parameters and save the modified centrifugation data to the same program location see Chapter Entering modifying programs As a result of saving the program linking will be removed Re link the programs see Chapter Linking programs 27 3 Centrifugation run with program linking Press the PROG key twice in order to select the parameter PR PART The ...

Page 19: ...materials or mixtures of materials with a higher density The permissible speed can be calculated using the following formula RPM speed maximum x kg dm density Greater 1 2 n speed Reduced red e g maximum speed RPM 4000 density 1 6 kg dm3 RPM 3464 RPM 4000 x kg dm 1 6 kg dm 1 2 nred In the exceptional case that the maximum loading indicated on the hanger is exceeded the speed must also be reduced Th...

Page 20: ...ge housing lid and centrifuging chamber 32 1 1 Surface cleaning and care Clean the centrifuge housing and the centrifuging chamber regularly using soap or a mild detergent and a damp cloth if required For one thing this services purposes of hygiene and it also prevents corrosion through adhering impurities Ingredients of suitable detergents soap anionic tensides non ionic tensides After using dete...

Page 21: ... the motor shaft should be lightly greased The rotors and the accessory parts have to be checked on a monthly basis for wear and corrosion For swing out rotors it is important to check the area of the lifting lugs for hangers the grooves and the base should be checked for cracks Rotors and attachments may no longer be utilised upon indication of wear and tear or corrosion Check the firm seating of...

Page 22: ...otors made of aluminium suspension made of metal lids with bio containment as well as stands and reductions can be autoclaved at 121 C 250 F 20 mins Otherwise you must ask the manufacturer No statement can be made about the degree of sterility The lids of the rotors and containers must be removed prior to autoclaving Autoclaving accelerates the ageing process of plastics In addition autoclaving ma...

Page 23: ...in switch position Ι The rotor must turn during switch on IMBALANCE UNWUCHT The rotor is unevenly loaded Open lid Check the loading of the rotor see chapter Loading the rotor Repeat the centrifugation run CONTROL ERROR 04 06 09 Error in lid locking or lid closure N MAX 05 Rotation too fast N MIN 13 Rotation too slow ROTORCODE 10 Incorrect rotor coding Perform a MAINS RESET MAINS INTERRUPT Power fa...

Page 24: ...evice must be decontaminated and cleaned to protect people the environment and property When you are disposing of the device the respective statutory rules must be observed Pursuant to guideline 2002 96 EC WEEE all devices supplied after August 13 2005 may not be disposed as part of domestic waste The device belongs to group 8 medical devices and is categorized in the business to business field Th...

Page 25: ...6 5243 4 5227 Röhrchen tube 2078 0536 0553 Ausschwingrotor 4 fach Swing out rotor 4 times 90 Kapazität capacity ml 1 5 2 0 12 50 5 Maße dimensions x L mm 11 x 38 17 x 100 29 x 115 12 x 75 Anzahl p Rotor number p rotor 64 28 8 80 Drehzahl speed RPM 4000 4000 4000 4000 RZB RCF 2826 2898 2755 2773 Radius radius mm 158 162 154 155 9 97 sec 42 9 sec 51 Temperatur temperature C 10 3 Probenerwärmung Samp...

Page 26: ...6 2 9 4 9 50 15 8 5 10 4 7 1 6 5 4 7 3 Maße dimensions x L mm 13 x 65 13 x 90 29 x 115 17 x 120 16 x 100 16 x 75 13 x 75 13 x 100 10 x 60 Anzahl p Rotor number p rotor 48 8 28 48 48 48 48 80 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 4000 4000 RZB RCF 2808 2844 2898 2755 2773 2808 2808 2737 Radius radius mm 157 159 162 154 155 157 157 153 9 97 sec 42 9 sec 51 Temperatur temperature C 10 3 Pr...

Page 27: ...71 1475 Ausschwingrotor 4 fach Swing out rotor 4 times 90 Kapazität capacity ml 4 8 3 x 2 4 x 1 1 x 8 2 x 8 Maße dimensions A mm 12 4 x 120 17 5 x 240 8 7 60 6 2 30 17 5 240 17 5 240 Anzahl p Rotor number p rotor 16 16 16 16 8 8 Filterkarten filter cards 1692 1691 1694 1693 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 RZB RCF 1735 2737 1735 2737 1735 2737 1735 2737 1664 2665 1664 2665 Radius r...

Page 28: ...ße dimensions x L mm 16 x 100 16 x 125 16 x 92 13 x 65 15 x 102 17 x 100 16 x 75 Anzahl p Rotor number p rotor 28 28 28 28 28 28 28 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 4000 RZB RCF 3005 3059 3005 3005 3005 3005 3005 Radius radius mm 168 171 168 168 168 168 168 9 97 sec 42 9 sec 51 Temperatur temperature C 10 3 Probenerwärmung Sample temp rise K 2 14 10 Tiefste erreichbare Temperatur b...

Page 29: ... Rotor number p rotor 32 32 32 32 32 32 32 32 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 4000 4000 RZB RCF 2952 2952 2952 2952 2952 2952 2952 2952 Radius radius mm 165 165 165 165 165 165 165 165 9 97 sec 42 9 sec 51 Temperatur temperature C 10 3 Probenerwärmung Sample temp rise K 2 14 10 Tiefste erreichbare Temperatur bei maximaler Drehzahl 1 h Laufzeit und 20 C Raumtemperatur nur bei Kühlz...

Page 30: ... 2 0 1 1 1 4 2 7 3 4 3 175 225 Maße dimensions x L mm 11 x38 8 x 66 11 x 66 10 x 88 10 x 60 61 x 118 61 x 137 Anzahl p Rotor number p rotor 40 40 40 40 40 4 4 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 RZB RCF 2916 2916 2916 2916 2916 3095 3095 Radius radius mm 163 163 163 163 163 173 173 9 97 sec 42 9 sec 51 Temperatur temperature C 10 3 Probenerwärmung Sample temp rise K 2 14 10 Tiefste er...

Page 31: ...s 90 Kapazität capacity ml 1 5 2 0 12 15 9 10 10 50 9 2 7 3 4 5 5 15 Maße dimensions x L mm 11 x 38 17 x 100 17 x 120 16x 92 15x 102 29 x 115 14 x 100 11 x66 11 x 92 17 x 100 Anzahl p Rotor number p rotor 96 42 66 12 72 120 120 72 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 4000 4000 RZB RCF 3363 3434 3291 3291 3309 3309 3309 3291 Radius radius mm 188 192 184 184 185 185 185 184 9 97 sec 45 9...

Page 32: ...6 2 30 8 7 60 12 4 120 17 5 240 8 7 60 6 2 30 6 2 30 8 7 60 Anzahl p Rotor number p rotor 12 12 12 12 12 12 24 24 Filterkarten filter cards 1675 1675 1675 1676 1677 1678 1692 1692 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 4000 4000 RZB RCF 2290 3291 2290 3291 2290 3291 2290 3291 2290 3291 2290 3291 2290 3291 2290 3291 Radius radius mm 128 184 128 184 128 184 128 184 128 184 128 184 128 184 ...

Page 33: ... dimensions x L mm 40 x 115 44 x 100 30 x 115 17 x 120 34 x 100 24 x 100 13 x 100 12 x 100 Anzahl p Rotor number p rotor 6 6 12 42 6 24 72 72 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 4000 4000 RZB RCF 3488 3488 3631 3631 3631 3434 3542 3542 Radius radius mm 195 195 203 203 195 192 198 198 9 97 sec 45 9 sec 55 Temperatur temperature C 10 0 Probenerwärmung Sample temp rise K 2 14 10 Tiefste ...

Page 34: ...17 x 120 17 x 100 16 x 75 17 x 100 Anzahl p Rotor number p rotor 42 42 42 42 42 42 42 48 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 4000 4000 RZB RCF 3542 3542 3542 3542 3542 3542 3542 3488 Radius radius mm 198 198 198 198 198 198 198 195 9 97 sec 45 9 sec 55 Temperatur temperature C 10 0 Probenerwärmung Sample temp rise K 2 14 10 Tiefste erreichbare Temperatur bei maximaler Drehzahl 1 h Lau...

Page 35: ...x 75 13 x 75 12 x 82 8 x 66 13 x 65 11 x 66 11 x 92 8 x 66 8 x 66 11 x 66 13 x 65 13 x 75 Anzahl p Rotor number p rotor 48 48 48 48 48 48 72 72 Drehzahl speed RPM 4000 4000 4000 4000 4000 4000 4000 4000 RZB RCF 3488 3488 3488 3488 3488 3488 3077 3077 Radius radius mm 195 195 195 195 195 195 172 172 9 97 sec 45 9 sec 55 Probenerwärmung Sample temp rise K 2 14 10 Tiefste erreichbare Temperatur bei m...

Reviews: